Large field-of-view short-wave infrared metalens for scanning fiber endoscopy

Ningzhi Xie, Matthew D. Carson, Johannes E. Fröch, Arka Majumdar, Eric J. Seibel, K. F. BöhringerJournal of Biomedical Optics 2023 · DOI

TL;DR

Researchers have developed a novel scanning fiber endoscope (SFE) that utilizes a metalens to significantly reduce its size while improving its optical performance, particularly at the edges of its field of view. This advancement holds promise for more effective cardiovascular disease diagnosis and surgery assistance by enabling clearer, wider-angle imaging within blood vessels.

What Problem Does This Paper Address?

Current scanning fiber endoscopes (SFEs) are crucial for minimally invasive procedures like cardiovascular diagnosis, offering a clear forward view into blood vessels. However, existing SFE systems, which rely on miniaturized refractive lens doublets, face limitations in terms of device length and off-axis optical aberrations. These limitations hinder further miniaturization and can degrade image quality at wider viewing angles. The short-wave infrared (SWIR) wavelength regime is particularly important for this application due to reduced scattering and water absorption, allowing visualization through blood without clearing. The challenge is to create an SWIR SFE that is not only smaller but also maintains or improves resolution and field of view (FOV) across a broad range of angles, especially when imaging deep within vessels.

Methods

The core innovation in this paper is the replacement of the traditional refractive lens doublet in an SFE with a transmissive metalens. The metalens was designed using Zemax Optical Studio, employing a binary2 phase mask optimized for a 1310 nm wavelength. This phase profile was then discretized into 12 levels and fabricated using electron beam lithography on a crystalline silicon-on-sapphire substrate. The meta-atoms (nanoposts) were designed to impart the desired phase response, with their dimensions theoretically calculated using rigorous-coupled wave analysis (RCWA). The performance of the metalens-based SFE was then characterized experimentally using a custom fiber scanner and a movable IR microscope. This involved measuring the longitudinal and transverse intensity distributions of the projected beam at various angles and distances to assess resolution, FOV, and depth of focus (DOF).

Datasets

Not discussed in the paper.

Key Insights and Ideas

  • Metalenses as a Game-Changer for Endoscopy: The paper demonstrates that metalenses can effectively replace bulky refractive optics in miniaturized imaging systems like SFEs, offering significant size reduction without compromising optical performance.
  • Aberration Correction at the Edge: A key advantage of the metalens is its superior performance at large off-axis angles compared to refractive lenses. This directly addresses the problem of resolution degradation at the edges of the FOV, which is critical for comprehensive imaging.
  • Optimized Design for Miniaturization and Performance: The design process involved a careful trade-off between reducing the optical track length (by decreasing the distance between the fiber tip and the metalens) and managing the beam's diverging angle, which impacts DOF and resolution.
  • SWIR Imaging Potential: The use of a metalens in the SWIR regime opens up possibilities for advanced biomedical imaging, allowing for better visualization in challenging environments like blood vessels.

Key Results

  • Device Miniaturization: The metalens reduced the optical track length of the SFE from 1.2 mm to 0.86 mm, a 28% reduction.
  • Large Field of View and Depth of Focus: The metalens-based SFE achieved a full FOV of 70 degrees and a DOF of approximately 15 mm.
  • Resolution: At the center of the field (15 mm illumination distance), the spatial resolution was approximately 140 µm.
  • Edge Performance: Resolution degradation at the edge of the FOV (35 degrees) was less than a factor of 2 for the metalens-based SFE, whereas the refractive lens counterpart experienced a ~3 times resolution degradation.
  • Aberration Reduction: The metalens demonstrated reduced spherical aberration at large projection angles compared to the refractive lens doublet.

Limitations and Future Work

The authors acknowledge that further device length reduction is possible by using thinner substrates and increasing the NA of the fiber. They also suggest that the current forward-view design could be adapted for side-view imaging by using a reflective metalens. Additionally, while this work focused on a single wavelength, the paper proposes that polychromatic metalenses could enable full-color visible SFEs, although this might require meta-atom engineering and could potentially lead to lower focusing efficiency. The authors also note that the measured beam longitudinal distribution slightly deviated from simulations, attributing this to the actual fiber beam deviating from a perfect point source model.

Summary updated

Abstract

These results show the promise of integrating a metalens into an endoscope for device minimization and optical performance improvement.