jULIEs: nanostructured polytrodes for low traumatic extracellular recordings and stimulation in the mammalian brain

Romeo Rácz, Mihály Köllő, Gabriella Racz, Ciprian Bulz, Tobias Ackels, Tom P. A. Warner, William Wray, Nikolai I. Kiskin, Chi Chen, Zhiwen Ye, Livia de Hoz, Ede Rancz, Andreas T. SchaeferJournal of Neural Engineering 2022 · DOI

TL;DR

Researchers have developed "jULIEs," ultra-low impedance, nanostructured glass-metal composite microwires for minimally invasive brain recording and stimulation. These probes offer high signal-to-noise ratios with minimal tissue damage, enabling detailed study of neural circuits in vivo.

What Problem Does This Paper Address?

Interrogating neural populations with extracellular microelectrodes is a cornerstone of neuroscience. However, a persistent challenge is the damage these probes inflict on brain vasculature and neural circuits during insertion, regardless of manufacturing method. This damage can compromise signal quality and long-term recording stability. The authors aim to overcome this by developing probes with an extremely small footprint, thereby minimizing tissue perturbation.

Methods

The core innovation lies in the fabrication and modification of "jULIEs" (just ultra-low impedance electrodes). These are made from continuously drawn SiO₂-insulated ultramicroelectrode fibers with diameters below 30 µm and conductive metal cores as small as 1 µm. To achieve low interfacial impedance, crucial for good signal coupling, a two-step electrochemical modification process is employed:

  1. Nano-gold (nanoAu) electrodeposition: This creates a rough, high-surface-area gold nanostructure on the exposed metal core.
  2. Iridium oxide (IrOx) electrodeposition: A film of polycrystalline IrOx is then deposited onto the nanoAu layer.

This nanostructuring significantly reduces interfacial impedance by up to two orders of magnitude, to below 500 kΩ at 1 kHz. The probes are then assembled into custom arrays.

Datasets

Not discussed in the paper.

Key Insights and Ideas

  • Minimally Invasive by Design: The extremely small diameter and smooth surface of the jULIE fibers, combined with their flexibility, allow for insertion with minimal disruption to blood vessels and neural tissue. Capillaries were observed to fold around the probes without rupture.
  • Electrochemical Nanostructuring is Key: The two-step nanoAu and IrOx deposition is critical for achieving low impedance and thus high signal-to-noise ratios, overcoming the inherent limitations of small electrode surface areas.
  • Scalable and Customizable Fabrication: The manufacturing process, from fiber drawing to electrochemical modification and array assembly, is described as scalable and adaptable to different neuroanatomical structures.
  • Robustness and Stability: The modified electrodes demonstrate mechanical stability during insertion and retraction, and their electrical properties remain stable even after extensive stimulation.
  • Simultaneous Recording and Stimulation: The low impedance and stable interface make jULIEs suitable for both high-quality recording and effective electrical stimulation.

Key Results

  • Low Interfacial Impedance: Achieved impedances below 500 kΩ at 1 kHz, a significant reduction from bare metal electrodes.
  • High Signal-to-Noise Ratio: Demonstrated exceptional signal-to-noise ratios for both single and multi-unit activity.
  • Minimal Tissue Damage: Histological and imaging studies showed no detectable damage to the blood-brain barrier or significant tissue displacement, unlike with conventional silicon probes.
  • Reliable Recordings: Unit activity was reliably resolved immediately after insertion, and recordings remained stable even after multiple insertion/retraction cycles.
  • Deep Brain Access: Successful recordings were achieved in deep brain structures down to >6 mm below the cortical surface.
  • Effective Stimulation: Cells responded to electrical stimulation with currents as low as 50 µA.

Limitations and Future Work

  • Acute vs. Chronic Implantation: The current study focuses on acute recordings. The authors suggest that the minimal tissue disturbance makes jULIEs ideal candidates for future long-term implantation studies.
  • Scalability to >10k Channels: While the technology is described as scalable, achieving very large-scale (above 10k) integration with standard readout electronics remains a challenge, though direct bonding and flip-chip bonding are mentioned as potential solutions.
  • Clinical Applications: The paper highlights the potential for clinical applications, but further development would be needed to translate these findings into clinical tools.

Summary updated

Abstract

<i>Objective.</i>Extracellular microelectrode techniques are the most widely used approach to interrogate neuronal populations. However, regardless of the manufacturing method used, damage to the vasculature and circuit function during probe insertion remains a concern. This issue can be mitigated by minimising the footprint of the probe used. Reducing the size of probes typically requires either a reduction in the number of channels present in the probe, or a reduction in the individual channel area. Both lead to less effective coupling between the probe and extracellular signals of interest.<i>Approach.</i>Here, we show that continuously drawn SiO<sub>2</sub>-insulated ultra-microelectrode fibres offer an attractive substrate to address these challenges. Individual fibres can be fabricated to >10 m continuous stretches and a selection of diameters below 30<i>µ</i>m with low resistance (<100 Ω mm<sup>-1</sup>) continuously conductive metal core of <10<i>µ</i>m and atomically flat smooth shank surfaces. To optimize the properties of the miniaturised electrode-tissue interface, we electrodeposit rough Au structures followed by ∼20 nm IrOx film resulting in the reduction of the interfacial impedance to <500 kΩ at 1 kHz.<i>Main results</i>. We demonstrate that these ultra-low impedance electrodes can record and stimulate both single and multi-unit activity with minimal tissue disturbance and exceptional signal-to-noise ratio in both superficial (∼40<i>µ</i>m) and deep (∼6 mm) structures of the mouse brain. Further, we show that sensor modifications are stable and probe manufacturing is reproducible.<i>Significance.</i>Minimally perturbing bidirectional neural interfacing can reveal circuit function in the mammalian brain<i>in vivo.</i>
BibTeX
@article{racz2022julies,
  author = {Romeo Rácz and Mihály Köllő and Gabriella Racz and Ciprian Bulz and Tobias Ackels and Tom P. A. Warner and William Wray and Nikolai I. Kiskin and Chi Chen and Zhiwen Ye and Livia de Hoz and Ede Rancz and Andreas T. Schaefer},
  title = {{jULIEs: nanostructured polytrodes for low traumatic extracellular recordings and stimulation in the mammalian brain}},
  journal = {Journal of Neural Engineering},
  year = {2022},
  doi = {10.1088/1741-2552/ac514f},
}
Figure