J164014 Design of a Rehabilitation Robot Suit with Mechanical Safety Devices : Detailed Design

Satoshi Kitaguchi, Yoshihiro Kai, Wenlong Zhang, Masayoshi TomizukaThe Proceedings of Mechanical Engineering Congress Japan 2013 · DOI

Abstract

Safety is one of the most critical challenges in the development of rehabilitation robot suits. This paper presents a knee-assist robot suit equipped with two hardware-based safety devices: a "velocity-based safety device" and a "torque-based safety device." These mechanisms consist entirely of mechanical components, operating independently of actuators, controllers, or batteries. This design ensures that safety is guaranteed even in the event of a computer or power failure. We review the design of these devices, present experimental results verifying their effectiveness, and describe the detailed integration into a rehabilitation robot suit.

Introduction: The Need for Passive Safety

In gait rehabilitation, robotic suits assist patients with neurological disorders in regaining mobility. However, the high power of these devices introduces risks; a software bug, sensor failure, or system crash could lead to excessive joint velocity or torque, potentially injuring the patient.

Most existing safety systems are "active," relying on the same electronic infrastructure that may be the source of the failure. To address this, the researchers developed "passive" mechanical safety devices that act as a physical fail-safe, operating autonomously from the robot's control logic.

The complete robot suit assembly and its application in a clinical setting.

Mechanical Safety Mechanisms

The robot suit features two distinct mechanical triggers designed to cut motor power when safe thresholds are exceeded. The core philosophy is that safety should be a physical law of the machine rather than a sub-routine in code.

1. Velocity-Based Safety Device

This device utilizes a rotary damper and a set of springs. During joint rotation, the damper generates resistance proportional to the angular velocity. When the velocity exceeds a pre-defined threshold, the damping torque overcomes the spring tension. This causes a mechanical bar to rotate and physically trigger a kill switch, immediately disconnecting the motor's power supply.

2. Torque-Based Safety Device

The torque-based mechanism monitors the physical displacement between two plates connected by a torsion spring. When the torque applied to the joint exceeds safe limits, the spring compresses, creating an angular difference between the plates. This displacement pulls a wire rope that slides a plate into a kill switch.

Exploded view of the robot suit showing the internal components of the velocity-based and torque-based safety devices.

A key advantage of these designs is adjustability. Thresholds for speed and force can be tuned by simply changing spring positions or component stiffness, requiring no software modifications.

Experimental Verification

To validate the precision of these mechanical triggers, trials were conducted using a prototype test rig equipped with a motion capture system (HAS-500) and a 6-axis force/torque sensor (IFS-100M40A).

  • Velocity Accuracy: For a target threshold of 0.88 rad/s, the device triggered at an average of 0.83 rad/s. At a higher setting of 2.18 rad/s, it triggered at 2.10 rad/s. The low standard deviation (0.11–0.21) demonstrates high repeatability.
  • Torque Accuracy: The torque device exhibited exceptional precision. For a 0.56 Nm target, the average trigger point was exactly 0.56 Nm with a standard deviation of only 0.015.
  • System Reliability: In all test cases, the mechanical triggers successfully cut motor power and halted joint movement immediately upon reaching the threshold.

Design Considerations and Future Work

The current design is optimized for human anatomy, with adjustable frames to fit thigh and calf lengths between 0.33m and 0.40m. While highly effective for a single joint (the knee), future work will explore integrating these mechanisms into multi-joint suits (hip and ankle).

Challenges remain regarding the manual reset required after a mechanical trip and the additional weight these components add to the suit. However, the primary goal—ensuring patient safety through physical laws rather than software routines—has been successfully demonstrated. The next phase involves full-scale clinical trials to evaluate the suit's performance in real-world rehabilitation scenarios.

Summary updated

Abstract

Safety is one of the most important issues in rehabilitation robot suits. We have presented a rehabilitation robot suit equipped with two hardware-based safety devices. The robot suit assists a knee joint of a patient. The safety devices consist of only mechanical components without actuators, controllers, or batteries. We call one device the "velocity-based safety device" and the other device the "torque-based safety device". We expect that the safety devices guarantee the safety even if the computer does not work. In this paper, we firstly review the rehabilitation robot suit equipped with the safety devices. Secondly, we present experimental results to verify the effectiveness of the prototype safety devices. Finally, we describe the design details of a rehabilitation robot taking into consideration the experimental results.
BibTeX
@inproceedings{kitaguchi2013j,
  author = {Satoshi Kitaguchi and Yoshihiro KAI and Wenlong Zhang and Masayoshi Tomizuka},
  title = {{J164014 Design of a Rehabilitation Robot Suit with Mechanical Safety Devices : Detailed Design}},
  booktitle = {The Proceedings of Mechanical Engineering Congress Japan},
  year = {2013},
  doi = {10.1299/jsmemecj.2013.\_j164014-1},
}
Figure