TL;DR
This paper investigates the effectiveness of three low-power millimeter-wave (mmWave) Frequency Modulated Continuous Wave (FMCW) radars operating at 24 GHz, 60 GHz, and 120 GHz for non-contact vital sign monitoring. The study demonstrates that higher frequencies and wider bandwidths lead to more accurate measurements of displacement and range, with the 120 GHz radar showing the best performance for estimating heart and respiratory rates in human subjects.
What Problem Does This Paper Address?
Monitoring vital signs like heart rate (HR) and respiratory rate (RR) is crucial for patient care, but current methods often require physical contact. This can be uncomfortable, intrusive, and pose a risk of infection, especially for vulnerable populations like infants or those with wounds. Non-contact methods are therefore highly desirable. While optical solutions exist, they are susceptible to environmental factors like lighting. Millimeter-wave (mmWave) radar technology offers a promising alternative for non-contact vital sign monitoring, capable of detecting the subtle chest movements associated with breathing and heartbeat. However, the optimal radar parameters, particularly carrier frequency and bandwidth, for achieving accurate and robust vital sign monitoring are not well-established. This paper aims to address this gap by systematically characterizing and comparing FMCW radars at different frequencies to determine their suitability for this application.
Methods
The researchers evaluated three low-power FMCW radar modules from Infineon Technologies: BGT24 (24 GHz), BGT60 (60 GHz), and BGT120 (120 GHz). These radars were configured with a sampling rate of 2 MHz and 128 samples per chirp, with chirps generated every 10ms.
The core methodology involved two main phases:
- Phantom Model Evaluation: A phantom model simulating human chest movement was created using a movable steel plate. This allowed for controlled testing of:
- Range Estimation: The accuracy of determining the distance to the target was assessed at various distances (30-60 cm).
- Baseline Noise: The variance of the phase signal for a static target was measured to quantify inherent noise levels, considering different angles of incidence (0°, 30°, 60°).
- Displacement Estimation: The phantom was set to oscillate at amplitudes mimicking heartbeat (0.08 mm, 0.3 mm) and breathing (1.2 mm) to evaluate the accuracy of displacement measurements.
- Human Subject Evaluation: Three male subjects were recorded for two minutes using each radar system. A previously published, low-complexity signal processing pipeline involving Fast Fourier Transform (FFT) was used to estimate HR and RR from the radar data. The ground truth for HR was obtained from an ECG sensor (Polar H10), and for RR, from an accelerometer on the same device.
Datasets
The paper does not explicitly mention training datasets. For evaluation, the authors used:
- Phantom Model Data: Data collected from a custom-built phantom that emulated human chest movements. This phantom was used to evaluate range, baseline noise, and displacement estimation accuracy. Two versions of the phantom were used: one with a bare metal plate and another covered with EMI-absorbing foam and a gelatinous layer to better emulate human body reflectance.
- Human Subject Data: Recordings from three human subjects performing vital sign monitoring. This data was used to evaluate the feasibility of estimating HR and RR with the radar systems and compare their performance against ground truth from a Polar H10 wearable device.
Key Insights and Ideas
- Frequency Matters: Higher radar frequencies (60 GHz and 120 GHz) significantly improve the accuracy of both range and displacement measurements compared to lower frequencies (24 GHz). This is attributed to the physics of wave propagation and the relationship between phase change and displacement, where higher frequencies yield larger phase shifts for the same displacement, making them easier to detect.
- Bandwidth Boosts Resolution: Wider chirp bandwidths, inherent in the higher frequency radars (e.g., 10 GHz for 120 GHz vs. 2 GHz for 24 GHz), directly translate to better range resolution. This finer resolution is crucial for distinguishing small movements.
- Phantom vs. Reality: While phantom models provide controlled environments for characterizing radar performance, the results from human subjects are essential for validating real-world applicability. The study successfully bridged this gap.
- Low-Power Viability: The investigation confirms that low-power mmWave FMCW radars can be effective for vital sign monitoring, paving the way for embedded, battery-operated devices.
- Algorithm Simplicity vs. Robustness: The paper highlights that while simple algorithms (like FFT) can identify vital signs, they can be prone to errors, such as mistaking breathing harmonics for heart rate. This suggests a need for more advanced algorithms for robust performance across diverse individuals and conditions.
Key Results
- Range Estimation: The 24 GHz radar showed a significant overestimation of distance (around 6 cm). The 60 GHz and 120 GHz radars achieved much lower errors (< 1 cm) with a metal plate, though the gelatin phantom increased variance.
- Baseline Noise: The 24 GHz radar exhibited higher baseline noise than the 60 GHz and 120 GHz radars across all angles and phantom types. The gelatin phantom generally showed slightly higher noise levels.
- Displacement Estimation: The BGT24 consistently showed larger errors in displacement estimation. The BGT60 and BGT120 performed similarly, with the BGT120 showing a slight advantage in the gelatin model for sub-millimeter displacements.
- Human Subject Vital Sign Estimation:
- HR: The 120 GHz radar achieved the lowest Mean Absolute Error (MAE) of 0.4 bpm with a standard deviation of 1 bpm. The BGT24 and BGT60 had MAEs of 4 bpm (σ=7) and 6 bpm (σ=7), respectively.
- RR: All radars performed similarly for RR estimation, with MAE < 1 breath per minute (brpm). The BGT24 showed some outliers for one subject.
Limitations and Future Work
- Sample Size and Variability: The study was conducted on only three human subjects. The authors acknowledge that a larger sample set is needed to develop algorithms that account for inter-subject variability (e.g., due to muscle, adiposity, breast tissue, or body hair).
- Algorithm Robustness: The current simple FFT-based algorithm for vital sign estimation can be prone to misinterpreting breathing harmonics as heart rate. The authors suggest that advanced algorithms are needed to tackle these issues.
- Distance Limitations: The current experiments were limited to a maximum distance of 60 cm. Future work could investigate optimizing radar systems for greater distances, which would be beneficial for different applications.
- Environmental Factors: While mmWave radar is less affected by lighting than optical methods, its performance in more complex environments (e.g., with significant clutter or movement) was not extensively explored.