Intra-Operative Optical Coherence Imaging of In-Vivo Chronic Otitis Media Followed by Post-Operative Audiogram Assessments

Ha Young Kim, Jaeyul Lee, Ruchire Eranga Wijesinghe, Jeong Hun Jang, Mansik Jeon, Jeehyun KimIEEE Journal of Selected Topics in Quantum Electronics 2020 · DOI

Abstract

The successful surgery of chronic otitis media (COM) is challenging; this depends on the surgeon's knowledge of the optical visibility of surgical microscopes. Herein, we reported the utilization of intra-surgical optical coherence tomography (OCT) system to effectively guide the surgery of COM based on augmented reality with cross-sectional images. The intra-surgical spectral-domain OCT system with a center wavelength of 846 nm was capable of obtaining non-invasive, high-resolution, and high-speed visualizations with an axial resolution of 8 μm, lateral resolution of 30 μm, and an extended working distance of 280 mm. Three patients with COM were involved in this research. The lesion conditions of the temporal bone were observed with computed tomography pre-operatively. Furthermore, pure-tone audiogram examinations were performed to evaluate pre and post-surgical conditions. The results revealed that the averaged air-bone gap of 500Hz, 1 kHz, 2 kHz, and 4kHz in all cases improved by 61%. Thus, the research proves that the experimental procedure can be beneficial and clinically applicable with the developed intra-surgical OCT system for future otolaryngological assessments.

Introduction

Chronic otitis media (COM) is a persistent inflammatory condition of the middle ear that often requires complex surgical intervention to restore hearing and prevent disease recurrence. Surgeons must navigate a delicate environment containing the tympanic membrane, ossicles, and the facial nerve. While traditional surgical microscopes provide excellent surface visibility, they cannot visualize subsurface pathology or the precise alignment of grafts. Pre-operative imaging, such as Computed Tomography (CT), provides a structural roadmap but lacks the real-time feedback and resolution (typically ~500 µm) necessary for intra-operative guidance. This study presents a custom intra-operative Optical Coherence Tomography (OCT) system integrated with a surgical microscope via Augmented Reality (AR) to provide high-resolution, real-time subsurface imaging during COM surgery.

System Architecture and Integration

The developed system utilizes a spectral-domain OCT (SD-OCT) with a center wavelength of 846 nm. A critical design feature is the extended working distance of 280 mm, achieved through customized magnification and objective lenses. This distance ensures that surgeons have sufficient space for physical instruments while maintaining continuous imaging.

Schematic of the intra-surgical SD-OCT system integrated with a surgical microscope, showing the OCT, microscope, and display beam paths.

The system features an AR integration where OCT cross-sections are projected directly into the surgeon's ocular eyepiece. This allows for simultaneous viewing of the surgical field and subsurface morphological data. High-speed processing is facilitated by a GPU-based architecture using CUDA kernels for real-time noise removal and Fourier transforms, achieving an axial resolution of 8 µm and a lateral resolution of 30 µm—significantly higher than conventional clinical imaging modalities.

Clinical Methodology and Insights

The clinical feasibility of the system was evaluated in three patients diagnosed with COM. Pre-operative assessments included CT scans and otoscopic examinations. During surgery, the AR-OCT system provided real-time volumetric scans and A-scan intensity profiles of the mastoid bone and tympanic membrane.

Key insights from the intra-operative use included:

  • Subsurface Visualization: The OCT identified residual inflammatory tissue and mastoid air cells invisible to the naked eye.
  • Quantitative Precision: The system allowed for micron-level measurement of membrane thickness and graft positioning, ensuring optimal placement.
  • Workflow Integration: A toggle feature allowed surgeons to activate the AR overlay only when necessary, preventing information overload.

Results and Discussion

Post-operative recovery was measured using pure-tone audiograms (PTA) three months after surgery. The clinical outcomes were robust, with the average air-bone gap (ABG) across the 500 Hz to 4 kHz range improving by 61%. All patients achieved hearing levels within the "serviceable" range (below 40 dB).

Visual confirmation of the AR-OCT system: (c-d) calibration on a phantom and (e-f) intra-operative view of the middle ear with the OCT cross-section overlay.

The 8 µm axial resolution of the OCT system provided a clear diagnostic advantage over the ~500 µm resolution of clinical CT, enabling the detection of minute bone erosions and thin membrane pathologies that are critical for successful surgical outcomes.

Conclusion

The integration of AR-OCT into middle ear surgery provides a powerful tool for real-time, high-resolution subsurface visualization. While the current study is limited by a small sample size and potential motion artifacts, the results demonstrate the clinical applicability of the system. Future work will focus on larger clinical trials and the expansion of this technology to other otolaryngological procedures, such as cochlear implantation and tumor resection near the facial nerve.

Summary updated

Abstract

The successful surgery of chronic otitis media (COM) is challenging; this depends on the surgeon's knowledge of the optical visibility of surgical microscopes. Herein, we reported the utilization of intra-surgical optical coherence tomography (OCT) system to effectively guide the surgery of COM based on augmented reality with cross-sectional images. The intra-surgical spectraldomain OCT system with a center wavelength of 846 nm was capable of obtaining non-invasive, high-resolution, and high-speed visualizations with an axial resolution of 8 μm, lateral resolution of 30 μm, and an extended working distance of 280 mm. Three patients with COM were involved in this research. The lesion conditions of the temporal bone were observed with computed tomography pre-operatively. Furthermore, pure-tone audiogram examinations were performed to evaluate pre and post-surgical conditions. The results revealed that the averaged air-bone gap of 500Hz,1 kHz, 2 kHz, and 4kHz in all cases improved to 61%. Thus, the research proves that the experimental procedure can be beneficial and clinically applicable with the developed intra-surgical OCT system for future otolaryngological assessments.
BibTeX
@article{kim2020intra,
  author = {Ha Young Kim and Jaeyul Lee and Ruchire Eranga Wijesinghe and Jeong Hun Jang and Mansik Jeon and Jeehyun Kim},
  title = {{Intra-Operative Optical Coherence Imaging of <i>In-Vivo</i> Chronic Otitis Media Followed by Post-Operative Audiogram Assessments}},
  journal = {IEEE Journal of Selected Topics in Quantum Electronics},
  year = {2020},
  doi = {10.1109/jstqe.2020.3022927},
}
Figure