Inter-scanner brain MRI volumetric biases persist even in a harmonized multi-subject study of multiple sclerosis

Kelly Clark, Carly M. O’Donnell, Mark A. Elliott, Shahamat Tauhid, Blake E. Dewey, Renxin Chu, Samar Khalil, Govind Nair, Pascal Sati, Anna DuVal, Nicole Pellegrini, Amit Bar‐Or, Clyde Markowitz, Matthew K. Schindler, Jonathan Zurawski, Peter A. Calabresi, Daniel S. Reich, Rohit Bakshi, Russell T. Shinohara, the Naims CooperativebioRxiv 2022 · DOI

TL;DR

Even with a standardized protocol, brain MRI scans from different scanners and sites show significant volumetric biases in multiple sclerosis (MS) patients. These biases persist across various automated analysis methods, highlighting the need for robust harmonization strategies in multi-center MS research and clinical practice.

What Problem Does This Paper Address?

Multicenter studies are crucial for recruiting enough participants to detect subtle effects in diseases like multiple sclerosis (MS). However, these studies often involve MRI scanners from different manufacturers and at various locations, leading to potential biases in the image data. These "site effects" or "scanner effects" can obscure true biological differences or treatment effects, making it difficult to compare results across sites or pool data. This paper investigates the extent of these inter-scanner and inter-site biases in brain volume measurements in MS patients, even when a harmonized imaging protocol is used.

Methods

The researchers employed a standardized, high-resolution 3-tesla MRI protocol across four different sites. They imaged 11 volunteers with stable MS, collecting both scan-rescan data at each site to assess intra-scanner variability. Several automated methods were used for brain volumetry, including:

  • MIMOSA: For automated white matter (WM) lesion segmentation.
  • Joint Label Fusion (JLF): For estimating normal-appearing gray matter (GM) and white matter (WM) volumes, as well as thalamic volumes.
  • FMRIB's Automated Segmentation Tool (FAST): For estimating whole brain volume.
  • FMRIB's Integrated Registration and Segmentation Tool (FIRST): For estimating thalamic volumes.

Manual lesion segmentation was also performed for comparison. Statistical analysis involved:

  • Random Effects Modeling: To assess differences in estimated volumes within and across sites.
  • Non-parametric Modeling: To understand the proportion of variation explained by site.
  • Permutation Testing: To rigorously test for significant differences in inter- and intra-scanner measurements, and to assess biases related to coil element activation.

Datasets

The study utilized data from 11 volunteers with stable multiple sclerosis (MS). These participants were scanned at four different sites, with nine participants scanned at all four sites. Two participants were scanned at fewer sites due to the COVID-19 pandemic. The data included both T1-weighted and T2-FLAIR MRI sequences.

Key Insights and Ideas

  • Persistence of Bias: The core finding is that even with a harmonized protocol, significant differences in brain volume measurements persist across different MRI scanners and sites. This suggests that scanner hardware, subtle protocol variations not fully captured by harmonization, or other site-specific factors contribute to these biases.
  • Site Explains Significant Variation: Non-parametric modeling revealed that "site" accounted for over 50% of the variation in most estimated volumes, and this increased to over 75% when data from different scanner manufacturers (Siemens and Philips) were included. This underscores the substantial impact of site on volumetric measures.
  • Inter-scanner vs. Intra-scanner Differences: Permutation tests showed significant differences between average inter-scanner and intra-scanner measurements for most brain volumes. This implies that the variability introduced by using different scanners is often greater than the variability within repeated scans on the same scanner.
  • Shading Artifacts Matter: An unexpected finding was the impact of spine coil activation due to head positioning. This created a shading artifact that led to significant differences in thalamic volume measurements, highlighting how even seemingly minor imaging artifacts can influence results.
  • Limitations of Standard Modeling: The study found that random effects models, commonly used in such analyses, violated assumptions of normal distribution and homogeneity of residuals. This led to an underestimation of the variance explained by site, suggesting that more advanced or specialized modeling techniques might be needed.

Key Results

  • Site explained > 50% of the variation in most estimated brain volumes across automated methods.
  • When data from both Siemens and Philips scanners were included, site explained > 75% of the variation.
  • Permutation tests revealed significant differences (P < .05) between average inter- and intra-scanner differences in most estimated brain volumes.
  • Inter-scanner differences were generally larger than intra-scanner differences, particularly when data from different scanner manufacturers were combined.
  • The activation of spine coil elements led to significant differences in thalamic volume measurements.

Limitations and Future Work

The authors acknowledge several limitations:

  • Limited Sample Size: The study involved only 11 participants and a limited range of scanners. This restricts the generalizability of the findings to a broader population, more scanner models, and different clinical settings (e.g., community hospitals).
  • Model Assumption Violations: The failure of standard random and mixed effects models due to non-normality and heteroskedasticity indicates a need for further development of appropriate parametric models to accurately represent site biases.
  • Artifact Impact: The shading artifact from spine coil activation highlights the sensitivity of volumetry to imaging artifacts, suggesting that further investigation into the effect of coil elements and scanning protocols is warranted.

Future work should focus on:

  • Larger, More Diverse Datasets: Including more participants, scanners, vendors, and versions to improve generalizability.
  • Longitudinal Studies: Investigating how site effects impact longitudinal changes in brain volume in MS patients.
  • Advanced Modeling: Developing and validating parametric models that can better account for site biases.
  • Mitigation Strategies: Exploring and testing methods to reduce or correct for inter-scanner and site-specific biases.

Summary updated

Abstract

Abstract Background/Purpose Multicenter study designs involving a variety of MRI scanners have become increasingly common. However, these present the issue of biases in image-based measures due to scanner or site differences. To assess these biases, we imaged 11 volunteers with multiple sclerosis (MS) with scan and rescan data at 4 sites. Materials and Methods Images were acquired on Siemens or Philips scanners at 3-tesla. Automated white matter lesion detection and whole brain, gray and white matter, and thalamic volumetry were performed, as well as expert manual delineations of T1 and T2 (FLAIR) lesions. Random effect and permutation-based nonparametric modeling was performed to assess differences in estimated volumes within and across sites. Results Random effect modeling demonstrated model assumption violations for most comparisons of interest. Non-parametric modeling indicated that site explained &gt; 50% of the variation for most estimated volumes. This expanded to &gt; 75% when data from both Siemens and Philips scanners were included. Permutation tests revealed significant differences between average inter- and intra-scanner differences in most estimated brain volumes ( P &lt; .05). The automatic activation of spine coil elements during some acquisitions resulted in a shading artifact in these images. Permutation tests revealed significant differences between thalamic volume measurements from acquisitions with and without this artifact. Conclusion Differences in brain volumetry persisted across MR scanners despite protocol harmonization. These differences were not well explained by variance component modeling; however, statistical innovations for mitigating inter-scanner differences show promise in reducing biases in multi-center studies of MS.
BibTeX
@misc{clark2022inter,
  author = {Kelly Clark and Carly M. O’Donnell and Mark A. Elliott and Shahamat Tauhid and Blake E. Dewey and Renxin Chu and Samar Khalil and Govind Nair and Pascal Sati and Anna DuVal and Nicole Pellegrini and Amit Bar‐Or and Clyde Markowitz and Matthew K. Schindler and Jonathan Zurawski and Peter A. Calabresi and Daniel S. Reich and Rohit Bakshi and Russell T. Shinohara and the NAIMS Cooperative},
  title = {{Inter-scanner brain MRI volumetric biases persist even in a harmonized multi-subject study of multiple sclerosis}},
  year = {2022},
  doi = {10.1101/2022.05.05.490645},
  howpublished = {bioRxiv},
}
Figure