Integrating medical physics into an EMR‐based radiology feedback system for quality improvement

Megan Russ, Justin Solomon, Steve Bache, Nicole M. Lafata, Erin B. Macdonald, Ehsan SameiJournal of Applied Clinical Medical Physics 2025 · DOI

TL;DR

This paper introduces a novel radiology quality feedback system integrated into an Electronic Medical Record (EMR) that actively involves medical physicists. By leveraging radiologist feedback, this system allows physicists to identify and address subtle image quality issues that might otherwise be missed, ultimately enhancing patient care and promoting continuous improvement in radiology practices.

What Problem Does This Paper Address?

Medical physicists play a crucial role in ensuring the quality and safety of medical imaging. However, their routine evaluations, often conducted annually or less frequently, are limited in scope and frequency. These evaluations might not capture the full spectrum of issues that arise during daily clinical practice, especially those that are subtle or specific to certain patient populations or imaging protocols. Radiologists, on the other hand, have a high sampling rate of patient images and are uniquely positioned to identify these nuanced quality issues. The challenge, therefore, is to effectively bridge this gap and create a system that allows radiologists to easily communicate these concerns to medical physicists, enabling timely and targeted interventions for quality improvement.

Methods

The researchers developed and implemented a new radiology quality feedback system within the Epic EMR system at a university hospital. This system is designed to be a seamless part of the radiologist's workflow, allowing them to submit feedback directly during report dictation via an optional, short survey. The survey includes checkboxes for various "quality detail" (e.g., noise, artifact, contrast) and "acquisition issues" (e.g., incorrect patient, protocol deviations), along with free-text fields for brief descriptions and positive feedback.

What's novel here is the explicit inclusion of medical physicists as recipients of this feedback, alongside technologist leaders. Submissions are routed to modality-specific teams comprising both physicists and technologists, who then investigate and address the reported issues. The paper focuses on evaluating the role and contributions of medical physicists within this system over a 31-month period.

Datasets

Not discussed in the paper. The study utilized feedback data collected through the implemented EMR system.

Key Insights and Ideas

  • Physicist-in-the-loop is key: Integrating medical physicists directly into the feedback loop, rather than just technologists, significantly enhances the ability to address technical quality issues.
  • Radiologists are the eyes and ears: Radiologists, with their high volume of image review, can identify subtle quality issues that routine QA might miss.
  • EMR integration is crucial for adoption: Embedding the feedback system within the existing EMR workflow (Epic) makes it accessible and encourages radiologist participation.
  • Beyond routine QA: This system uncovers issues that fall outside the scope of traditional, scheduled quality assurance procedures.
  • Collaborative problem-solving: The system fosters collaboration between radiologists, technologists, and physicists, leading to more comprehensive solutions.

Key Results

Over a 31-month period, 515 follow-up tickets were generated from radiologist feedback. Medical physicists responded to 9.3% of these tickets, while technologists handled the remaining 90.7%. Physicists were more heavily involved in resolving "quality detail" issues (28.3% of these tickets) compared to "acquisition issue" tickets (5.8%). Specifically, physicists were instrumental in addressing issues related to artifacts, noise, and image contrast.

The paper highlights three illustrative examples:

  1. Radiography: Correcting image processing settings to optimize grayscale range for AP chest X-rays, resolving excessive contrast issues.
  2. CT: Investigating and resolving image noise in chest/abdomen/pelvis scans by adjusting CTDIvol parameters, leading to a discussion about standardizing scan protocols.
  3. Ultrasound: Analyzing ticket trends revealed a high percentage of incomplete exams due to missing views/measurements, prompting protocol revisions and staff education.

Limitations and Future Work

The authors acknowledge that while technologists handled the majority of tickets, physicists' involvement was crucial for specific technical quality issues. Some messages were excluded from the analysis, including those with only positive feedback and tickets from modalities still being integrated into the system (fluoroscopy, mammography, nuclear medicine).

Future work could involve expanding the system to include these other modalities and further exploring the impact of this feedback system on patient outcomes and overall diagnostic accuracy. The increasing ubiquity of EMR systems like Epic suggests that similar feedback tools could be implemented widely to improve radiology practices globally.

Summary updated

Abstract

This work demonstrates the value of radiology quality feedback systems and the opportunity to address issues not typically identified during routine medical physics quality assurance. By leveraging radiologist feedback, physicists can enhance clinical practice, promote continuous improvement, and ensure consistent, high-quality imaging and safety for patients.
BibTeX
@article{russ2025integrating,
  author = {Megan Russ and Justin Solomon and Steve Bache and Nicole M. Lafata and Erin B. Macdonald and Ehsan Samei},
  title = {{Integrating medical physics into an EMR‐based radiology feedback system for quality improvement}},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70227},
}

Loading related content...

Figure