Integrated Piezoelectric Vibration and In Situ Force Sensing for Low-Trauma Tissue Penetration

Bingsheng He, Yao Guo, Guang‐Zhong YangCyborg 2025 · DOI

TL;DR

This paper introduces an innovative integrated piezoelectric module (IPEM) that combines high-frequency vibration with in-situ force sensing for minimally invasive tissue penetration. By using a single piezoelectric element for both actuation and sensing, the IPEM significantly reduces tissue trauma and insertion forces, paving the way for smarter surgical instruments.

What Problem Does This Paper Address?

Precise insertion of micro-scale instruments into delicate biological tissues, such as during neural probe implantation or ophthalmic surgery, is crucial for minimizing damage. Current methods often rely on rapid insertions to overcome tissue resistance, which can lead to inflammation and trauma. Furthermore, existing tools often lack real-time force feedback, making it difficult to adapt to tissue variations and ensure procedural safety. The challenge lies in developing miniaturized, integrated systems that can both assist penetration and provide accurate force measurements without adding bulk or complexity.

Methods

The core innovation is the Integrated Piezoelectric Module (IPEM). This module cleverly utilizes the dual nature of piezoelectric materials:

  • Vibration-Assisted Penetration: The inverse piezoelectric effect is used to generate high-frequency axial vibrations (around 4,652 Hz) in a tungsten probe. This vibration helps to break down tissue resistance and facilitate smoother penetration.
  • In-Situ Force Sensing: The direct piezoelectric effect is employed to measure the electrical charge generated by mechanical stress on the same piezoelectric element. This allows for real-time, localized force sensing at the probe tip without the need for external sensors.

The IPEM consists of two piezoelectric discs (one for actuation, one for sensing) stacked together, with a tungsten probe rigidly attached. The system's signal flow involves a digital-to-analog converter (DAC) and power amplifier for driving the actuator, and a charge amplifier, low-pass filter, and analog-to-digital converter (ADC) for processing the sensor's output.

Datasets

Not discussed in the paper. The experiments were conducted using gelatin phantoms and in vivo mouse brains.

Key Insights and Ideas

  • Single Element, Dual Function: The most significant insight is the successful integration of both actuation (vibration) and sensing (force) into a single, compact piezoelectric component. This drastically simplifies the instrument's design and reduces its footprint.
  • Vibration as a "Lubricant": The high-frequency vibrations don't just break tissue; they appear to dynamically "lubricate" the probe-tissue interface, reducing friction and adhesion, which is crucial for both insertion and retraction.
  • Real-time Feedback is Key: The ability to sense forces in real-time during vibration-assisted insertion opens the door for intelligent control algorithms that can adapt to tissue heterogeneity and prevent unexpected damage.
  • Mimicking Surgical Procedures: The system's ability to accurately replicate the implantation and detachment sequence of a neural electrode in vivo demonstrates its practical relevance for advanced surgical tasks.

Key Results

  • Force Sensing Accuracy: The piezoelectric sensor demonstrated excellent linearity (R² = 0.9998) and sensitivity (9.3 mV/mN) in static calibration. Dynamic tests showed mean absolute errors (MAE) below 0.3 mN and mean absolute percentage errors (MAPE) below 1% under various conditions, comparable to commercial sensors.
  • Vibration-Assisted Force Reduction:
    • In gelatin phantoms, peak puncture force was reduced by approximately 13%.
  • In vivo mouse brain experiments, the peak insertion force was reduced by approximately 33% (from 11.67 mN to 7.8 mN) when vibration was activated.
  • Thermal Safety: Temperature increases during vibration were minimal (0.5 °C in the IPEM, 0.1 °C in the phantom), well below the neural damage threshold.
  • Vibration Performance: The system achieved a dominant axial vibration amplitude of ±8.8 µm at its resonant frequency of ~4.6 kHz, with minimal lateral displacement, which is ideal for precise penetration.

Limitations and Future Work

  • Electrical Noise Sensitivity: The system can be susceptible to electrical noise, which might require advanced filtering in noisy surgical environments.
  • Environmental Factors: Temperature and humidity variations can affect piezoelectric material performance, potentially impacting accuracy.
  • Tissue Heterogeneity: While the system shows promise, the current tests were on relatively uniform phantoms and specific in vivo conditions. Variability in tissue properties could lead to unpredicted force spikes.
  • Response Time: While adequate for many procedures, the response time might lag in ultrafast applications.

The authors propose several exciting future directions:

  • Closed-Loop Control: Integrating real-time force feedback to modulate vibration amplitude/frequency, adjust insertion speed/trajectory, and anticipate tissue rupture events.
  • Mechanical Compatibility: Investigating and tuning the system for different tissue types (e.g., cornea, tumors) with varying mechanical properties.
  • Surgical Robotics Integration: Addressing challenges like sterilization compatibility, rapid probe exchange, and further noise suppression for clinical use.
  • Multichannel/Array Implantation: Extending the system for implanting multiple electrodes simultaneously.

Summary updated

Abstract

Precision-controlled microscale manipulation tasks-including neural probe implantation, ophthalmic surgery, and cell membrane puncture-often involve minimally invasive membrane penetration techniques with real-time force feedback to minimize tissue trauma. This imposes rigorous design requirements on the corresponding miniaturized instruments with robotic assistance. This paper proposes an integrated piezoelectric module (IPEM) that combines high-frequency vibration-assisted penetration with real-time in situ force sensing. The IPEM features a compact piezoelectric actuator integrated with a central tungsten probe, generating axial micro-vibration (4,652 Hz) to enable smooth tissue penetration while simultaneously measuring contact and penetration forces via the piezoelectric effect. Extensive experiments were conducted to validate the effectiveness and efficacy of the proposed IPEM. Both static and dynamic force-sensing tests demonstrate the linearity, sensitivity (9.3 mV/mN), and accuracy (mean absolute error < 0.3 mN, mean absolute percentage error < 1%) of the embedded sensing unit. In gelatin phantom tests, the module reduced puncture and insertion forces upon activation of vibration. In vivo experiments in mouse brains further confirmed that the system could reduce penetration resistance (from an average of 11.67 mN without vibration to 7.8 mN with vibration, decreased by 33%) through the pia mater and accurately mimic the electrode implantation-detachment sequence, leaving a flexible electrode embedded with minimal trauma. This work establishes a new paradigm for smart surgical instruments by integrating a compact actuator-sensor design with real-time in situ force feedback capabilities, with immediate applications in brain-machine interfaces and microsurgical robotics.
BibTeX