Instrumented shoulder functional assessment using inertial measurement units for frozen shoulder

Ting-Yang Lu, Kai-Chun Liu, Chia-Yeh Hsieh, Chih-Ya Chang, Yu Tsao, Chia-Tai ChanarXiv 2021 · DOI · arXiv

Abstract

Frozen shoulder (FS) is a condition characterized by pain and a significant loss of shoulder range of motion, hindering daily independence. While traditional assessments rely on manual measurements, this study proposes an objective, instrumented approach using inertial measurement units (IMUs). By analyzing kinematic parameters—such as smoothness, power, speed, and duration—across both complete tasks and specific subtasks, the research demonstrates that smoothness features effectively differentiate FS patients from healthy controls ($p < 0.05$). Notably, subtask analysis reveals subtle clinical information that is often obscured when evaluating tasks in their entirety.

Introduction

Frozen shoulder (FS), or adhesive capsulitis, severely restricts a person's ability to perform basic daily activities. Current clinical assessments typically use manual tools to measure Range of Motion (ROM) or rely on subjective patient questionnaires. However, manual measurements are prone to human error, and questionnaires can be influenced by subjective bias. More importantly, simple ROM measurements fail to capture the quality of movement or the specific phases of a complex task where a patient struggles. This work explores the use of wearable sensors to provide a more nuanced, objective assessment of shoulder function.

Methods

The study recruited 40 participants: 20 FS patients and 20 healthy controls. Each participant was equipped with two IMUs—one on the wrist and one on the upper arm. They performed five functional tasks representative of daily life: Washing hair (WH), Washing upper back (WUB), Washing lower back (WLB), Placing an object on a high shelf (POH), and Removing an object from a back pocket (ROP).

A novel aspect of this methodology is the manual segmentation of each task into three distinct subtasks: initiation/reach, the main activity, and the return to the starting position. This allows for a granular analysis of movement quality throughout the entire motion cycle.

Table 1: Detailed breakdown of the five functional tasks and their corresponding subtasks.

Researchers extracted several kinematic features, including smoothness (measured via "jerk" and acceleration peaks), power (Power Index), speed (Range of Angular Velocity), and duration.

Key Results and Discussion

The results indicate that movement smoothness is a critical biomarker for FS. Patients often exhibit "stuttering" or less fluent movements as they compensate for pain or stiffness. The "Number of Peaks in Acceleration" (NP-A) was particularly effective at highlighting these differences between groups.

Figure 1: Comparison of sensor data (acceleration and angular velocity) between a healthy subject and an FS patient, illustrating the differences in signal smoothness and subtask duration.

The Importance of Subtasks

One of the study's major findings is that subtask analysis reveals "hidden" deficits. For example, in the "Removing object from back pocket" (ROP) task, the total duration might appear normal, but the initial rotation (Subtask 1) often shows significant impairment in FS patients. Similarly, tasks like washing the lower back (WLB) and reaching for a high shelf (POH) were the most challenging, showing significant differences in power and speed across nearly all subtasks.

Conclusion

This IMU-based assessment provides a more comprehensive view of functional impairment than traditional ROM measurements. By focusing on movement quality and subtask performance, clinicians can better identify specific functional limitations and track recovery progress more accurately. Future work will focus on automating the subtask segmentation process and correlating these objective metrics with the clinical stages of frozen shoulder.

Summary updated

Abstract

Frozen shoulder (FS) is a shoulder condition that leads to pain and loss of shoulder range of motion. FS patients have difficulties in independently performing daily activities. Inertial measurement units (IMUs) have been developed to objectively measure upper limb range of motion (ROM) and shoulder function. In this work, we propose an IMU-based shoulder functional task assessment with kinematic parameters (e.g., smoothness, power, speed, and duration) in FS patients and analyze the functional performance on complete shoulder tasks and subtasks. Twenty FS patients and twenty healthy subjects were recruited in this study. Five shoulder functional tasks are performed by participants, such as washing hair (WH), washing upper back (WUB), washing lower back (WLB), placing an object on a high shelf (POH), and removing an object from back pocket (ROP). The results demonstrate that the used smoothness features can reflect the differences of movement fluency between FS patients and healthy controls (p < 0.05 and effect size > 0.8). Moreover, features of subtasks provided subtle information related to clinical conditions that have not been revealed in features of a complete task, especially the defined subtask 1 and 2 of each task.
BibTeX
@misc{lu2021instrumented,
  author = {Ting-Yang Lu and Kai-Chun Liu and Chia-Yeh Hsieh and Chih-Ya Chang and Yu Tsao and Chia-Tai Chan},
  title = {{Instrumented shoulder functional assessment using inertial measurement units for frozen shoulder}},
  year = {2021},
  eprint = {2111.13312},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2111.13312},
}
Figure