Influence of the Area of the Aqueduct on Quantification of Stroke Volume and Max Velocity in Healthy Volunteers Using Phase Contrast Cine MRI

Hongri Zhang, Weike Duan, Xiaopan Li, Yixin Wang, Xinyu Li, Petrice M. Cogswell, Benjamin D. ElderResearch Square 2021

Abstract

Background: The relationship of the area of the aqueduct on quantification of the aqueductal stroke volume (SV) and max velocity need further investigation. Our aim was to assess the influence of the area of the aqueduct on quantification of the aqueductal SV and max velocity measured with phase contrast magnetic resonance imaging (PC-MRI) within the cerebral aqueduct at the level of the intercollicular sulcus.

Materials and Methods: Nine healthy volunteers (mean age 29.6 yrs) were enrolled in the study and brain MRIs were performed on a 3.0T system. Quantitative analysis of aqueductal cerebrospinal fluid (CSF) flow was performed using manual regions of interest (ROI) placement. ROIs were separately drawn for each of 12 phases of the cardiac cycle, and changes in aqueduct size during the cardiac cycle were determined. Stroke volumes were calculated using the first and ninth aqueductal ROIs and compared to each other. Max velocities at the 12 phases were also collected, and the relationship between the area and max velocity and the impact on SV were analyzed.

Results: There was variation in the size of the aqueduct during the cardiac cycle; the first area ($S1$) was larger than the ninth ($S9$). The first max velocity ($V_{max1}$) was less than the ninth ($V_{max9}$). Additionally, there was a significant difference between the stroke volume calculated using the first aqueductal ROI ($SV1$) and the ninth ($SV9$).

Conclusions: There is variation in the size of the cerebral aqueduct which is used to calculate stroke volume and other CSF flow parameters during the cardiac cycle. The maximum velocity may be inversely proportional to the area of the aqueduct. In order to establish reliable reference values for CSF flow parameters in future studies, a variable ROI, to account for cardiac cycle variation, should be considered and incorporated.

Introduction

Quantifying cerebrospinal fluid (CSF) flow through the cerebral aqueduct is a cornerstone of diagnosing and managing neurological conditions such as hydrocephalus and Chiari malformations. The gold standard for this assessment is Phase Contrast Cine MRI (PC-MRI), which captures fluid velocity over the course of a cardiac cycle.

However, most current post-processing techniques rely on a "static ROI." A radiologist typically draws a single boundary around the aqueduct on one image frame and applies that same boundary to all other frames in the cardiac cycle to calculate flow parameters. This study questions a fundamental assumption: Does the aqueduct stay the same size throughout a heartbeat? If the area changes, a static ROI might include surrounding brain tissue or exclude part of the fluid, leading to inaccurate data.

Methods

The researchers studied nine healthy volunteers (mean age 29.6) using a 3.0T MRI scanner. They employed a standard PC-MRI sequence, but with a high-effort manual analysis:

  • Phase Gating: The cardiac cycle was divided into 12 distinct phases.
  • Manual Segmentation: Instead of drawing one ROI, radiologists manually drew ROIs for each of the 12 phases for every participant.
  • Comparative Analysis: They specifically compared the first phase ($S1$) and the ninth phase ($S9$) of the cycle, as these showed the most significant differences in area.
  • Recalculation: They calculated Stroke Volume (SV) and Maximum Velocity ($V_{max}$) using these different phase-specific ROIs to see how much the "area factor" influenced the final numbers.

Results

The study revealed that the cerebral aqueduct is not a static structure; its area fluctuates significantly during a single heartbeat. The area at the first phase ($S1$) was significantly larger than at the ninth phase ($S9$).

Table 1: Maximum velocity measurements across the cardiac cycle for nine subjects.

As shown in Table 1, the maximum velocity ($V_{max}$) varied across the cardiac cycle phases. The maximum velocity at phase 1 ($V_{max1}$) was significantly lower than at phase 9 ($V_{max9}$), with values of approximately -1.16 cm/s vs -5.77 cm/s ($P = 0.001$). This confirms an inverse proportionality: as the aqueduct narrows, the fluid velocity increases.

Table 2: Quantitative variation in aqueductal parameters across the cardiac cycle for nine subjects.

Furthermore, recalculating stroke volume using different ROIs yielded significantly different results. $SV1$ ($0.0044 \pm 0.0034$ ml) was significantly lower than $SV9$ ($0.0061 \pm 0.0033$ ml) ($P = 0.0282$). These discrepancies, illustrated by the data in Table 2, highlight the measurement error introduced by static ROIs.

Discussion

The most critical insight is that the aqueduct's area is dynamic. Brain tissue expands during arterial systole (when blood rushes in), which physically squeezes the aqueduct, making it smaller. During diastole, the tissue shrinks, and the aqueduct expands. Because stroke volume is calculated as Mean Velocity × Area, using an ROI from the "wrong" phase of the heartbeat can significantly skew results. If a large ROI from phase 1 is used to calculate flow in phase 9 (where the aqueduct is actually smaller), measurement error is inevitable.

Conclusion

The study concludes that a "variable ROI" that adjusts to the aqueduct's size in each phase is necessary for reliable CSF flow reference values. While manually drawing 12 ROIs per patient is not feasible in a clinical setting, future work should focus on developing automated, variable ROI segmentation software to implement these findings practically. This would improve the accuracy of neuroimaging diagnostics for conditions involving CSF dynamics.

Summary updated

BibTeX
@inproceedings{zhang2021influence,
  author = {Hongri Zhang and Weike Duan and Xiaopan Li and Yixin Wang and Xinyu Li and Petrice M. Cogswell and Benjamin D. Elder},
  title = {{Influence of the Area of the Aqueduct on Quantification of Stroke Volume and Max Velocity in Healthy Volunteers Using Phase Contrast Cine MRI}},
  booktitle = {Research Square},
  year = {2021},
  doi = {10.21203/rs.3.rs-334057/v1},
}
Figure