Increasing the sensor channels: a solution for the pressing offsets that cause the physiological parameter inaccuracy in radial artery pulse signal acquisition

Chao Chen, Zhendong Chen, Hongmiin Luo, Bo Peng, Yinan Hao, Xiaohua Xie, Haiqing Xie, Xinxin LiFrontiers in Bioengineering and Biotechnology 2024 · DOI

Abstract

Introduction: In studies of pulse wave analysis, single-channel sensors only adopt single temporal pulse signals without spatial information to show pulse-feeling patterns. Multi-channel arterial pulse signals, also named as three-dimensional pulse images (3DPIs), provide the spatial and temporal characteristics of radial pulse signals. When involving single or few-channel sensors, pressing offsets have substantial impacts on obtaining inaccurate physiological parameters like tidal peak ($P_2$). Methods: This study discovers the pressing offsets in multi-channel pulse signals and analyzes the relationship between the pressing offsets and time of $P_2$ ($T_2$) by qualifying the pressing offsets. First, we employ a data acquisition system to capture 3DPIs. Subsequently, the $errorT_2$ is developed to qualify the pressing offsets. Results: The outcomes display a central low and peripheral high pattern. Additionally, the $errorT_2$ increase as the distances from the artery increase, particularly at the radial ends of the blood flow direction. For every 1 mm increase in distances between sensing elements and center sensing elements, the $errorT_2$ in the radial direction escalates by 4.87%. When the distance is greater than 3.42 mm, the $errorT_2$ experiences a sudden increase. Discussion: The results show that increasing the sensor channels can overcome the pressing offsets in radial pulse signal acquisition.

Introduction

Pulse Wave Analysis (PWA) serves as a vital non-invasive tool in cardiovascular diagnostics, enabling the detection of conditions such as arterial stiffness and endothelial dysfunction. A key parameter in this analysis is the tidal peak ($P_2$), which represents the second wave peak within a cardiac cycle. However, the accuracy of PWA is often compromised by "pressing offsets"—measurement errors that occur when a sensor is not perfectly aligned with the radial artery.

Traditional single-channel sensors capture only temporal data, lacking the spatial context necessary to identify if a signal is distorted by misalignment. This study explores how multi-channel tactile sensors can provide three-dimensional pulse images (3DPIs) to characterize and mitigate these pressing offsets.

Methodology

To address the limitations of single-point sensing, the researchers utilized a multi-channel data acquisition system.

Hardware and Data Collection

The system featured a Pressure Profile Systems (PPS) tactile sensor with a $5 \times 5$ grid of sensing elements, providing 24 active channels. This array allows for the simultaneous capture of pulse signals across a spatial area, creating a 3DPI that reflects both temporal and spatial characteristics of the radial pulse.

Figure 6: Sensor placement on the radial artery and the resulting 3D and 2D pulse signal amplitude distributions.

Signal Processing and Quantification

Raw data was processed using a novel de-noising algorithm: Cross Channel Dynamic Weighting Principal Component Analysis (cc-DWRPCA). This method leverages correlations between adjacent channels to effectively remove motion artifacts and noise.

To quantify the impact of pressing offsets, the researchers developed the $errorT_2$ metric. This measures the time discrepancy of the $P_2$ peak between a peripheral sensing element and the central element positioned directly over the artery. By aggregating these errors across 52 subjects (623 samples), an error Stacked Surface (error$T_2$SS) was generated to map spatial inaccuracy.

Key Results and Insights

The study revealed several critical findings regarding how spatial misalignment affects pulse signal quality:

  • Spatial Error Gradient: The $errorT_2$ follows a "central low, peripheral high" pattern. The timing of the tidal peak is most stable directly over the artery and degrades predictably as the sensor moves toward the periphery.
  • Directional Sensitivity: Measurement error is significantly more sensitive to offsets in the radial direction (along the blood flow) compared to the axial direction. Specifically, for every 1 mm increase in distance from the center in the radial direction, the $errorT_2$ increases by approximately 4.87%.
  • The 3.42 mm Threshold: A critical discovery was the identification of a quality "cliff." When the pressing offset exceeds 3.42 mm, the $errorT_2$ undergoes a sudden, sharp increase, indicating that data beyond this point is highly unreliable for physiological parameter extraction.

Discussion and Conclusion

The findings demonstrate that increasing the number of sensor channels is a robust solution for the pressing offset problem. By providing spatial redundancy, multi-channel arrays ensure that even if the sensor is not perfectly placed, at least one channel will likely fall within the low-error zone (under the 3.42 mm threshold).

This spatial information allows for "self-correcting" diagnostics, where the system can automatically identify the optimal sensing channel. Future work will focus on expanding the demographic diversity of the study and integrating these insights into real-time clinical feedback systems to improve the reliability of cardiovascular monitoring.

Summary updated

Abstract

<b>Introduction:</b> In studies of pulse wave analysis, single-channel sensors only adopt single temporal pulse signals without spatial information to show pulse-feeling patterns. Multi-channel arterial pulse signals, also named as three-dimensional pulse images (3DPIs), provide the spatial and temporal characteristics of radial pulse signals. When involving single or few-channel sensors, pressing offsets have substantial impacts on obtaining inaccurate physiological parameters like tidal peak (P<sub>2</sub>). <b>Methods:</b> This study discovers the pressing offsets in multi-channel pulse signals and analyzes the relationship between the pressing offsets and time of P2 (T<sub>2</sub>) by qualifying the pressing offsets. First, we employ a data acquisition system to capture 3DPIs. Subsequently, the errorT<sub>2</sub> is developed to qualify the pressing offsets. <b>Results:</b> The outcomes display a central low and peripheral high pattern. Additionally, the errorT<sub>2</sub> increase as the distances from the artery increase, particularly at the radial ends of the blood flow direction. For every 1 mm increase in distances between sensing elements and center sensing elements, the errorT<sub>2</sub> in the radial direction escalates by 4.87%. When the distance is greater than 3.42 mm, the errorT<sub>2</sub> experiences a sudden increase. <b>Discussion:</b> The results show that increasing the sensor channels can overcome the pressing offsets in radial pulse signal acquisition.
BibTeX
@inproceedings{chen2024increasing,
  author = {Chao Chen and Zhendong Chen and Hongmiin Luo and Bo Peng and Yinan Hao and Xiaohua Xie and Haiqing Xie and Xinxin Li},
  title = {{Increasing the sensor channels: a solution for the pressing offsets that cause the physiological parameter inaccuracy in radial artery pulse signal acquisition}},
  booktitle = {Frontiers in Bioengineering and Biotechnology},
  year = {2024},
  doi = {10.3389/fbioe.2024.1359297},
}
Figure