NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

In-vivo left atrial surface motion and strain measurement using novel mesh regularized image block matching method with 4D-CTA

Han Yu, Zidun Wang, Hao Wu, Zhengduo Zhu, Jiaqiu Wang, Runxing Fang, Shanglin Wu, Hujin Xie, Xianjue Huang, Jessica Benitez Mendieta, Haveena Anbananthan, Zhiyong LiJournal of Biomechanics 2024

Abstract

Atrial strain and motion play important roles in evaluation of stroke risks for patients with atrial fibrillation. While cardiac computed tomographic angiography (CTA) provides detailed left atrial morphology with unparallel image resolution, finding a suitable strain measurement method for CTA remains a considerable challenge. In this paper, for the first time, we introduced a mesh regularized image block matching method to estimate 3D left atrial (LA) surface strain with 4D CTA. A series of performance tests with ex-vivo phantom and in-vivo 4D-CTA data were deployed. In conclusion, our proposed method could provide reliable LA motion and strain data within limited time.
BibTeX
@article{yu2024vivo,
  author = {Han Yu and Zidun Wang and Hao Wu and Zhengduo Zhu and Jiaqiu Wang and Runxing Fang and Shanglin Wu and Hujin Xie and Xianjue Huang and Jessica Benitez Mendieta and Haveena Anbananthan and Zhiyong Li},
  title = {{In-vivo left atrial surface motion and strain measurement using novel mesh regularized image block matching method with 4D-CTA}},
  journal = {Journal of Biomechanics},
  year = {2024},
  doi = {10.1016/j.jbiomech.2024.112354},
}

Loading related content...

Figure