In vitro study of coronary flow occlusion in transcatheter aortic valve implantation.

Zhengfu He, Weiming Zhang, G Lutter, René Quaden, Jochen Cremer, Xiujun CaiPubMed 2014 · DOI

TL;DR

This study investigates the anatomical relationship between the aortic root, coronary ostia, and aortic valve leaflets in postmortem human hearts to understand why transcatheter aortic valve implantation (TAVI) can obstruct coronary blood flow. They found that a significant majority of coronary ostia are positioned below the sinotubular junction (STJ) and are thus susceptible to being covered by native leaflets after stent implantation, highlighting the need for improved TAVI stent designs.

What Problem Does This Paper Address?

Transcatheter aortic valve implantation (TAVI) has emerged as a life-saving alternative for patients with severe aortic stenosis who are too frail for traditional open-heart surgery. However, TAVI is not without its risks, and one serious complication is the obstruction of coronary arteries. This can occur because the implanted valve stent, or the native aortic leaflets pushed by the stent, can block the openings (ostia) that supply blood to the heart muscle itself. While this complication is known, the precise anatomical reasons for its occurrence, particularly the relationship between the coronary ostia, the aortic valve leaflets, and the sinotubular junction (STJ) – a key anatomical landmark in the aortic root – were not fully elucidated. This paper aims to fill that gap by providing detailed measurements of these structures in postmortem hearts to better understand the potential for coronary occlusion during TAVI.

Methods

The researchers conducted an in vitro study using 40 postmortem human hearts. They meticulously dissected the aortic roots and performed precise measurements of various anatomical features. Key measurements included:

  • The width of the aortic leaflets.
  • The height of the aortic sinus annulus to the sinotubular junction (STJ).
  • The distance from the aortic sinus annulus to the coronary ostia.
  • The distance from the coronary ostia to the STJ level.

Crucially, they simulated the implantation of a valved stent in the aortic annulus position. This allowed them to observe and evaluate the relationship between the implanted stent, the native aortic leaflets, and the coronary ostia both before and after stent deployment, and after opening the aorta. They specifically looked at whether the native leaflets covered the coronary ostia after stent implantation.

Datasets

The study utilized 40 postmortem human hearts obtained from autopsies at the Department of Pathology of University Hospital of Schleswig-Holstein, Germany. These hearts belonged to white adults of both genders. No specific datasets were created or used for training; the study relied on direct anatomical measurements from these specimens.

Key Insights and Ideas

  • Coronary Ostia Location is Key: The most striking finding is that a substantial majority (nearly three-quarters) of coronary ostia are located below the sinotubular junction (STJ). This anatomical positioning makes them vulnerable to being covered by the aortic leaflets.
  • Leaflet Coverage is Common: Following the simulated stent implantation, the native aortic leaflets were observed to fold upwards and become immobilized against the aortic wall. This often resulted in the partial or full covering of the coronary ostia by these leaflets.
  • Anatomical Basis for Complication: The study provides a clear anatomical explanation for why coronary occlusion can occur during TAVI. The proximity of the coronary ostia to the STJ and their relationship with the aortic leaflets are critical factors.
  • Need for Stent Modification: The findings strongly suggest that current TAVI stent designs may not adequately account for the anatomical variations in coronary ostia location, leading to potential obstruction. This underscores the need for the development of modified stents that can prevent coronary flow occlusion.

Key Results

  • Approximately 72.5% of the coronary ostia were found to be located below the STJ level.
  • In 55% of the postmortem specimens, one or both coronary ostia were partially or fully covered by the native aortic leaflets after simulated stent implantation.
  • When coronary ostia were located below the STJ, they were covered by leaflets in 72.4% of cases. In contrast, ostia located at or above the STJ were covered much less frequently (7.7%).
  • The study found that the right coronary ostia were, on average, located slightly lower (deeper) within the valsalva sinus than the left coronary ostia.

Limitations and Future Work

The authors acknowledge several limitations:

  • Specimen Age and Condition: The average age of the postmortem specimens was 69.7 years, which is younger than the typical TAVI patient population (around 82 years). Furthermore, the aortic leaflets and annulus tissues were normal or mildly calcified, whereas TAVI patients often have severe aortic stenosis and significant calcification, which could alter anatomical relationships.
  • In Vitro vs. In Vivo: The study was conducted in vitro using postmortem hearts. A dynamic, physiological environment in vivo might yield different results.
  • Stent Type: The study used a specific type of valved stent, and future research should explore different stent designs and implantation techniques to compare their impact on coronary flow.

The authors suggest that future work should focus on designing new TAVI stents that specifically address the issue of coronary ostia occlusion. They also emphasize the importance of thorough pre-procedural anatomical evaluation using imaging techniques like echocardiography, CT, or MRI to identify patients at higher risk for coronary obstruction and to inform surgical planning.

Summary updated

Abstract

Most of the coronary ostia were located below the STJ level and could be covered by the leaflets. This highlights the need of modified stents to prevent occlusion of coronary flow after TAVI.
BibTeX
@inproceedings{he2014vitro,
  author = {Zhengfu He and Weiming Zhang and G Lutter and René Quaden and Jochen Cremer and Xiujun Cai},
  title = {{In vitro study of coronary flow occlusion in transcatheter aortic valve implantation.}},
  booktitle = {PubMed},
  year = {2014},
  doi = {10.3978/j.issn.2072-1439.2014.12.19},
}