In-bore Preamplifier Nonlinearity Calibration

Chris Vassos, John M. Pauly, Fraser Robb, Shreyas Vasanawala, Greig ScottProceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition 2024

Abstract

Present MRI preamplifiers are very linear but use 10s of mA bias current and over 150mW power. For wireless arrays, low power alternatives based upon SiGe transistors are needed. These achieve the needed gain and noise figure, with order of magnitude power reduction, but at the cost of increased nonlinearity. Using pilot tone technology, we present a calibration system for preamplifier nonlinearity correction targeted to low power SiGe preamplifiers. This system integrated a phase-locked SDR into a form of prescan to demonstrate that in-bore preamplifier linearity calibration and correction are possible.