Improved Contrast-Enhanced Ultrasound Imaging Based on Microbubble Blind Deconvolution
Abstract
Ultrasound contrast harmonic imaging enhances image quality by employing inverse Wiener-filter deconvolution of the ultrasound contrast agents (UCAs) point spread function (PSF) and second-harmonic signals to suppress tissue echo components; however, current methods relying on simulated microbubble signals from computational models yield suboptimal PSF estimation accuracy, consequently compromising imaging performance. To address this limitation, this study proposes a Deconvolution approach based on Real UCAs PSF (DRUP) for improved contrast harmonic imaging, wherein inverse Wiener-filter deconvolution effectively recovers the original second-harmonic signal while suppressing tissue echoes. In vivo experiments on UCAs-perfused rabbit kidneys demonstrated that DRUP significantly outperforms simulated UCAs methods, achieving a 51.58% increase in contrast-to-tissue ratio, a 32.13% increase in contrast-to-noise ratio, and a 22.86% improvement in image contrast. Collectively, this method generates higher quality UCAs-enhanced images and delivers more precise diagnostic information for assessing deep-tissue microvascular perfusion.