Implementation and evaluation of the half-scan scheme based on CBCT (cone-beam CT) system

Dong Yang, Ruola Ning, Yong Yu, David L. Conover, Xianghua LuProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2004

Abstract

A modified Feldkamp (FDK) 3-D cone beam algorithm was developed to conduct the half-scan scheme on a flat panel detector (FPD)-based prototype CBCT system which is available in our Lab. X-ray source scans the object in a circular trajectory in the range of 180 degrees plus full cone angle rather than 360 degrees for a full scan. The redundant data is weighted by a weighting function in 3-D case based on the ones proposed by Parker1, FDK algorithm is then adopted to get the reconstruction image. The breast phantom and breast specimen are used in the experiment and the objects reconstructed from half-scan CBCT are compared with those reconstructed from full scan. The applicability of the half-scan scheme is testified in terms of image noise level, contrast resolution. The half-scan scheme is expected to improve the temporal resolution and may also reduce the patients’ X-ray dose level. The result showed an encouraging potential usage of the Half-scan CBCT in the functional 4-D CT imaging.