Impact of filter configurations on bipolar EGMs: An optimal filter setting for identifying VT substrates

Masateru Takigawa, Frédéric Sacher, Claire Martin, Ghassen Cheniti, Josselin Duchâteau, Thomas Pambrun, Nicolas Derval, Hubert Cochet, Mélèze Hocini, Tasuku Yamamoto, Takuro Nishimura, Susumu Tao, Shinsuke Miyazaki, Masahiko Goya, Tetsuo Sasano, Michel Haissaguierre, Pierre Jaı̈sJournal of Cardiovascular Electrophysiology 2023 · DOI

Abstract

Bipolar EGMs are strongly affected by filter settings in scar/border areas. In all, 30-250 or 30-500 Hz may be the best configuration, minimizing the baseline fluctuation, baseline noise, and detecting LAVAs. Not applying the 50-Hz notch filter may be beneficial to avoid missing VT substrate.
BibTeX
@article{takigawa2023impact,
  author = {Masateru Takigawa and Frédéric Sacher and Claire Martin and Ghassen Cheniti and Josselin Duchâteau and Thomas Pambrun and Nicolas Derval and Hubert Cochet and Mélèze Hocini and Tasuku Yamamoto and Takuro Nishimura and Susumu Tao and Shinsuke Miyazaki and Masahiko Goya and Tetsuo Sasano and Michel Haissaguierre and Pierre Jaı̈s},
  title = {{Impact of filter configurations on bipolar EGMs: An optimal filter setting for identifying VT substrates}},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2023},
  doi = {10.1111/jce.15997},
}