Image-based Motion Correction for the Siemens hybrid-MR/BrainPET Scanner

J. Scheins, Cr Brambilla, Jörg Mauler, Elena Rota Kops, Lutz Tellmann, Christoph Lerche, N. Jon ShahNuklearmedizin 2019 · DOI

Abstract

Head motion can degrade accuracy of time activity curves (TAC) in quantitative neuroimaging; this becomes especially true for long PET acquisition protocols. Apart from misalignment of regions of interest, also a bias occurs due to mismatches between the subject attenuation map (AM) in a fixed reference position and the various head positions. Any mismatch causes a bias of the attenuation correction (AC), but also of the scatter correction (SC) which usually depends on the AM. Thus, motion correction (MC) is a prerequisite for quantification. Where no external tracking device is available, motion parameters can be estimated directly from PET images. We have implemented a robust MC workflow based on PET image rigid co-registration. Now, it can be routinely used for neuroimaging studies with our Siemens hybrid MR/BrainPET scanner [1].