ICP-MS-based competitive immunoassay for the determination of total thyroxin in human serum
Abstract
To further expand the range of analytes that can be detected by using ICP-MS coupled with bioanalytical methods, we have employed a new separation system based on the highly active surface streptavidin and biotinylated monoclonal antibody (McAb) in a competitive immunoassay followed by ICP-MS detection. Specifically, we have demonstrated its application for the determination of total thyroxine (T4) in human serum using Eu3+ as a label. In this method, streptavidin immobilized to the pre-coated bovine serum albumin (BSA)-biotin on the microwells showed a significantly higher binding capacity for the biotinylated anti-T4 monoclonal antibody. Total T4 was quantified by measuring the Eu3+ intensity with ICP-MS after using 1% HNO3 to extract Eu3+ from the bound fraction of the immunocomplex. The detection limit was 7.4 ng mL−1 with a sample volume of 25 µL. Total T4 values obtained by this procedure agreed well with those obtained by a chemiluminescent immunoassay method. Both inter- and intra-assay precisions were below 10%. The results indicate that a general competitive ICP-MS-based immunoassay scheme may be possible.
BibTeX
@article{zhang2002icp,
author = {Chao Zhang and Fengbo Wu and Xinrong Zhang},
title = {{ICP-MS-based competitive immunoassay for the determination of total thyroxin in human serum}},
journal = {Journal of Analytical Atomic Spectrometry},
year = {2002},
doi = {10.1039/b205623b},
}Ask Althea to put this paper to the test.