High precision capacitance displacement measurement circuit design

Xuefeng Xing, Jianqiang Li, Shixue Lv, Linlin Xu, Xiaopu Zhang2015

Abstract

A high-resolution digital integrated chip capacitor measurement system was designed. The major components of the system include a main control chip, capacitance transducer, data storage, and communication interface. An ARM7-based microcontroller, the STM32F207, was used as the main control chip, while the AD7747 capacitance-to-digital converter, featuring a sensitivity of 20 aF, served as the transducer. The system supports a capacitance measurement range of -8 pF to +8 pF and a common-mode immutable capacitance up to 17 pF. Following systematic error correction derived from fixed capacity testing, the final measurement error was reduced to less than 3%.

Introduction

In the fields of precision machining and industrial automation, the demand for ultra-precision displacement sensors has grown significantly. Capacitive sensors are particularly advantageous due to their mechanical simplicity, high-temperature stability, and non-contact operation. By using air as a dielectric, these sensors avoid the internal friction and self-heating issues common in resistive or optical alternatives. However, the extremely low signal levels—often in the attofarad range—require specialized circuitry to convert minute capacitance changes into precise digital data without degradation from noise.

System Architecture and Methods

The researchers developed a digital integrated system architecture designed to maintain high resolution while simplifying the analog front-end.

System architecture block diagram showing the integration of the STM32F207 microcontroller and AD7747 transducer.

The system is built around two primary integrated circuits:

  1. AD7747 Transducer: This high-resolution capacitance-to-digital converter provides a sensitivity of 20 aF ($10^{-18}$ Farads). It handles the conversion of capacitance changes from a unipolar plate sensor into a digital signal, effectively replacing complex discrete analog bridge circuits.
  2. STM32F207 Microcontroller: Serving as the system's brain, this ARM7-based controller manages the I2C communication with the AD7747, handles data storage via FLASH and SDRAM, and facilitates PC communication through a UART interface.

To ensure accuracy, the team implemented a software-based calibration process. By performing a "fixed capacitor test" with known standards, they identified a consistent systematic absolute error of 0.222 pF. Subtracting this offset from raw readings significantly improved the final measurement precision.

Experimental Results and Analysis

The system's performance was characterized using a series of standard capacitors ranging from 3.3 pF to 15.0 pF. The experimental data was used to validate the system's linearity and precision.

Experimental data comparing standard capacitance values with measured values to determine system error.

Key technical specifications achieved include:

  • Sensitivity: 20 aF.
  • Linearity and Precision: 0.01% and 10 fF, respectively.
  • Measurement Range: -8 pF to +8 pF, with common-mode support up to 17 pF.
  • Update Rate: 5 Hz to 45 Hz.
  • Final Accuracy: After correcting for the mean absolute error, the final measurement error was brought down to less than 3%.

Conclusion

This design demonstrates that integrating a specialized capacitance-to-digital converter with a powerful microcontroller can achieve high-precision displacement tracking with minimal hardware complexity. While the system performs exceptionally for unipolar sensors, future work will address the narrow linear range and saturation issues encountered when using differential capacitance plates.

Summary updated

Abstract

A high resolution digital integrated chip capacitor measurement system was designed.The major components of the system mainly includedmain control chip, capacitance transducer, data storage and communication interface.ARM7-based mcro controllerSTM32F207was used as the main control chip, capacitance measurement chip AD7747produced by AD Company with sensitivity can be up to 20 af was used as capacitance transducer. The capacitance measurement range ofAD7747 is -8 pf to +8pf, the maximum common-mode immutable capacitance can reach17 pf anddata update rate range is between 5 Hz to 45 Hz. The absolute error of test circuit was calculated by the fixed capacity test and error analysis. The final results were obtained after subtracting the absolute error, and the final test error is less than 3%.
BibTeX
@misc{xing2015high,
  author = {Xuefeng Xing and Jianqiang Li and Shixue Lv and Linlin Xu and Xiaopu Zhang},
  title = {{High precision capacitance displacement measurement circuit design}},
  year = {2015},
  doi = {10.2991/ic3me-15.2015.206},
}
Figure