High-Performance Limiting Current Oxygen Sensor Comprised of Highly Active La0.75Sr0.25Cr0.5Mn0.5O3 Electrode

Jie Zou, Qian Lin, Chu Cheng, Xin Zhang, Qinghui Jin, Han Jin, Jinxia Wang, Jiawen Jian2018 · DOI

TL;DR

This paper introduces a promising alternative to expensive platinum electrodes in limiting current oxygen sensors. By utilizing a highly active perovskite material, La$_{0.75}$Sr$_{0.25}$Cr$_{0.5}$Mn$_{0.5}$O$_{3}$ (LSCM), the researchers demonstrate a sensor with comparable performance to platinum-based sensors, including fast response times, linear output, and excellent selectivity, paving the way for more cost-effective oxygen sensing applications.

What Problem Does This Paper Address?

Efficient combustion of fossil fuels is crucial for maximizing energy conversion and minimizing harmful exhaust emissions. Achieving optimal combustion relies heavily on precise control of the air-to-fuel (A/F) ratio, which in turn necessitates accurate oxygen sensing. Limiting current oxygen sensors are particularly attractive due to their stable signals and wide monitoring range. However, the widespread adoption of these sensors has been hindered by the high cost of the platinum electrodes traditionally used. This paper addresses the need for a cost-effective yet high-performance alternative to platinum in limiting current oxygen sensors.

Methods

The researchers synthesized La$_{0.75}$Sr$_{0.25}$Cr$_{0.5}$Mn$_{0.5}$O$_{3}$ (LSCM) powder using a sol-gel method, followed by calcination at temperatures up to 1400 °C to ensure a stable perovskite crystal phase. They then fabricated both LSCM-based and platinum-based limiting current oxygen sensors using a high-temperature co-firing technology (HTCC) with an 8YSZ electrolyte.

Key techniques employed for characterization and evaluation include:

  • X-ray Diffraction (XRD): To confirm the crystal structure and phase purity of the synthesized LSCM powders.
  • Scanning Electron Microscopy (SEM): To examine the microstructure of the LSCM electrode and the sensor's cross-section, including particle size and interface formation.
  • Energy Dispersive Spectroscopy (EDS): To analyze the elemental composition and confirm the absence of solid solution formation between LSCM and YSZ.
  • Electrochemical Impedance Spectroscopy (EIS): To evaluate the electrocatalytic activity of the LSCM and Pt electrodes by analyzing the interfacial resistance at various temperatures.
  • Linear Sweep Voltammetry (LSV): To obtain current-voltage (I-V) characteristics and identify the limiting current plateau.
  • Gas Sensing Tests: To assess the sensor's response and recovery times, linearity, selectivity against CO$_{2}$ and H$_{2}$O, and overall performance under varying oxygen concentrations.

Datasets

Not discussed in the paper. The study focuses on the fabrication and testing of novel sensor devices rather than training machine learning models on existing datasets.

Key Insights and Ideas

  • LSCM as a Viable Platinum Alternative: The core insight is that LSCM exhibits electrocatalytic activity towards oxygen reduction comparable to that of platinum, making it a strong candidate for replacing expensive Pt electrodes in limiting current oxygen sensors.
  • Thermal Stability of LSCM: LSCM maintains its perovskite crystal phase even after thermal treatment at a high temperature of 1400 °C, which is crucial for the co-firing process with the 8YSZ electrolyte.
  • Interface Formation: The study highlights the formation of a satisfactory contact between the porous LSCM electrode and the dense 8YSZ electrolyte, which is essential for efficient electrochemical reactions.
  • Electrocatalytic Activity Equivalence: EIS measurements reveal that the low-frequency impedance, representing interfacial catalytic activity, shows similar trends for both LSCM and Pt electrodes as a function of temperature, supporting the idea of equivalent catalytic performance.
  • Identical Sensing Performance: Remarkably, the LSCM-based sensors demonstrate sensing characteristics (response/recovery time, linearity, selectivity) that are virtually identical to those of the Pt-based sensors, validating the hypothesis of LSCM's suitability.
  • Negligible Hysteresis: The LSCM-based sensors exhibit negligible hysteresis, particularly at lower bias voltages, which is attributed to minimal polarization effects.

Key Results

  • Fast Response/Recovery: Both LSCM and Pt sensors achieved quick response and recovery times, typically within 7 seconds, even at operational temperatures as low as 420 °C.
  • Linear Response: A near-perfect linear relationship was observed between the sensor's response signal and oxygen concentration in the range of 2-25%, with a correlation coefficient of 0.999 for both LSCM and Pt sensors.
  • High Selectivity: The sensors demonstrated excellent selectivity to oxygen, with minimal response to 5% CO$_{2}$ and H$_{2}$O.
  • Comparable Limiting Currents: At 20% oxygen and 580 °C, the limiting currents for LSCM and Pt sensors were very close (24.029 μA vs. 24.009 μA), with a relative error of less than 0.08%.
  • Stable LSCM Phase: XRD patterns confirmed that LSCM calcined at temperatures up to 1400 °C retained its pure perovskite crystal phase.

Limitations and Future Work

  • Solid Solution Formation: While EDS confirmed no solid solution formation between LSCM and YSZ after thermal treatment, the paper notes that LSCM can only form an "imitate contact" with dense 8YSZ at temperatures above 1400 °C, suggesting potential for further optimization of interface engineering.
  • Hysteresis at Higher Voltages: Minor hysteresis was observed for both LSCM and Pt sensors at bias voltages below 0.5 V, though this was attributed to tiny polarization and considered negligible. Further investigation into minimizing this at higher voltages might be beneficial.
  • Long-Term Stability: While the study demonstrates comparable performance, long-term operational stability and durability of the LSCM-based sensors under various harsh conditions were not extensively evaluated. Future work could focus on assessing the long-term performance and reliability of these sensors.
  • Broader Application Range: The study focused on oxygen concentrations from 2-25%. Expanding the tested concentration range and evaluating performance under different atmospheric conditions would be valuable for real-world applications.

Summary updated

BibTeX
@misc{zou2018high,
  author = {Jie Zou and Qian Lin and Chu Cheng and Xin Zhang and Qinghui Jin and Han Jin and Jinxia Wang and Jiawen Jian},
  title = {{High-Performance Limiting Current Oxygen Sensor Comprised of Highly Active La0.75Sr0.25Cr0.5Mn0.5O3 Electrode}},
  year = {2018},
  doi = {10.3390/s18072155},
}
Figure