Abstract
Hydrogen-deuterium exchange (HDX) experiments analyzed by mass spectrometry (MS) provide information about the dynamics and the solvent accessibility of protein backbone amide hydrogen atoms. Continuous improvement of MS instrumentation has contributed to the increasing popularity of this method; however, comprehensive automated data analysis is only beginning to mature. We present Hexicon 2, an automated pipeline for data analysis and visualization based on the previously published program Hexicon (Lou et al. 2010). Hexicon 2 employs the sensitive NITPICK peak detection algorithm of its predecessor in a divide-and-conquer strategy and adds new features, such as chromatogram alignment and improved peptide sequence assignment. The unique feature of deuteration distribution estimation was retained in Hexicon 2 and improved using an iterative deconvolution algorithm that is robust even to noisy data. In addition, Hexicon 2 provides a data browser that facilitates quality control and provides convenient access to common data visualization tasks. Analysis of a benchmark dataset demonstrates superior performance of Hexicon 2 compared with its predecessor in terms of deuteration centroid recovery and deuteration distribution estimation. Hexicon 2 greatly reduces data analysis time compared with manual analysis, whereas the increased number of peptides provides redundant coverage of the entire protein sequence. Hexicon 2 is a standalone application available free of charge under http://hx2.mpimf-heidelberg.mpg.de.
Introduction
Hydrogen-Deuterium Exchange Mass Spectrometry (HDX-MS) is a powerful technique for "watching" proteins in motion, revealing how their structures shift and interact with ligands by measuring how quickly hydrogen atoms on the protein backbone swap with deuterium in the surrounding solvent. However, the data produced is massive: a series of 2D liquid chromatography-mass spectrometry (LC-MS) maps across multiple time points.
Traditionally, analyzing this data has been a manual "bottleneck." Researchers had to painstakingly assign peptide sequences, align retention times across samples, and calculate mass shifts. Existing software often struggled with noisy data, required instrument-specific formats, or produced "over-sparse" results that didn't accurately reflect the true distribution of deuterium. Hexicon 2 aims to provide a robust, instrument-independent, and fully automated solution to these challenges.
Methods
Hexicon 2 builds upon the original Hexicon pipeline but introduces several critical algorithmic upgrades to handle high-resolution data and improve accuracy.
The workflow consists of several key stages:
- Divide-and-Conquer Peak Detection: It uses the NITPICK algorithm for de novo feature detection but applies it in segments to handle high-resolution data without the quadratic runtime penalties of the original version.
- Nonlinear Chromatogram Alignment: To match peptides across different experimental runs, the software uses a piecewise linear regression. This allows it to predict elution times even when the chromatography shifts significantly between samples.
- Internal Recalibration: High-resolution MS data can suffer from calibration drift. Hexicon 2 uses dominant, well-identified mass peaks to internally recalibrate the spectra, significantly reducing false discovery rates in sequence assignment.
- Gold’s Iterative Deconvolution: This is perhaps the most significant upgrade. While the original Hexicon used a "sparse" regression that often missed subtle distribution details, Hexicon 2 employs Gold’s algorithm (borrowed from nuclear physics). This method is much more robust to noise and provides smoother, more realistic estimations of how deuterium is distributed across a peptide population.
- Integrated Result Browser: A new GUI allows researchers to visually inspect data, perform quality control, and export publication-quality figures.
Key Insights and Results
The authors benchmarked Hexicon 2 using the AppA–PpsR system from Rhodobacter sphaeroides, a complex biological system involving protein-protein interactions and light-dependent conformational changes. The evaluation included seven different conditions (free proteins vs. complexes, dark vs. light states), totaling dozens of LC-MS maps.
The Necessity of Recalibration
One of the most critical findings was the impact of mass accuracy on sequence assignment. The authors demonstrated that even a small calibration error (e.g., 16 ppm) can lead to a 50% failure rate in sequence assignment. The internal recalibration feature effectively "rescues" these datasets, maintaining high performance even in the presence of drift.
Performance Metrics
- Full Coverage: Hexicon 2 achieved 100% sequence coverage for both AppA and PpsR, outperforming manual analysis which yielded 91-97%.
- High Accuracy: Automated mass centroid extraction deviated from manual "gold standard" extraction by a median of only 0.12 Da (roughly 3% of total deuteration).
- Efficiency: The software processed a complete experimental series in less than one working day. Individual datasets took between 5 to 8 minutes to process on a standard desktop PC.
- Robustness: 95% of the automated results differed by less than 22% from manual analysis, even in complex, overlapping regions of the spectra.
- Redundancy as Quality Control: By detecting peptides de novo, Hexicon 2 achieves high redundancy (averaging 15 features per amide), allowing for higher confidence in final structural interpretations.
Conclusion
Hexicon 2 represents a major step forward in automating HDX-MS data analysis. By reducing the analysis time from weeks to hours while maintaining or exceeding manual accuracy, it enables more high-throughput studies of protein dynamics. While challenges remain—particularly with highly overlapping peaks—the current iteration provides a robust, redundant, and user-friendly platform for the scientific community. Future work may involve implementing mixture models to better deconvolute overlapping signals.
Summary updated