Halo performance on low light level image intensifiers
Lists
Only you see this
Abstract
To analyze the formation mechanism of the halo on low light level image intensifiers and the influencing factors on the halo size, a halo tester has been designed. Under the illumination between 10<sup>-2</sup> lx and 10<sup>-4</sup> lx, we use the tester to collect a 0.1922 mm hole image directly with CoolSNAP<sub>K4</sub> charge-coupled device (CCD) in a darkroom. The practical measurement result shows that the amplification ratio is 343.4. Then we put the super second and third generation image intensifiers after the hole, and the halo sizes of the hole images on the screens are determined as 0.2388 and 0.5533 mm respectively. The results are helpful to improve the quality of the low light level image intensifiers.
BibTeX
@inproceedings{cui2013halo,
author = {Dongxu Cui and Ling Ren and Benkang Chang and Feng Shi and Jifang Shi and Yunsheng Qian and Honggang Wang and Junju Zhang},
title = {{Halo performance on low light level image intensifiers}},
booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
year = {2013},
doi = {10.1117/12.2019621},
}Loading related content...