Abstract
For volunteers, the performance of T2WI as well as DWI with three respiratory-trigger techniques was similarly good. As for patients, FT-DWI is superior to BT and NT techniques in terms of higher lesion SNR and CNR at b50.
Introduction
Abdominal MRI is inherently challenging due to respiratory motion, which often leads to image artifacts like ghosting and blurring. These artifacts can obscure small lesions, complicating diagnosis. Current clinical practice relies on three primary strategies: breath-holding, which is difficult for compromised patients; respiratory belts (BT), which require physical contact and setup time; and navigator-triggers (NT), which track the diaphragm but can be inefficient. This study evaluates a novel, contactless frequency-modulated continuous-wave (FMCW) radar-trigger (FT) technique designed to improve both workflow efficiency and patient comfort without compromising diagnostic quality.
Methods
This prospective, single-center trial involved 40 subjects, including 17 healthy volunteers and 23 patients with liver diseases (primarily liver cancer and hemangiomas). Imaging was performed on a 1.5T MRI scanner integrated with a 4 GHz FMCW radar system. The radar system automatically reconstructs respiratory curves from body movement without requiring physical contact. The study compared the FT technique against the industry-standard BT and NT methods. Evaluation included qualitative assessments by radiologists using Likert scales and quantitative measurements of Signal-to-Noise Ratio (SNR), Contrast-to-Noise Ratio (CNR), and Apparent Diffusion Coefficient (ADC) values.
Results and Discussion
Workflow Efficiency and Scan Time
The FT technique significantly streamlines the MRI workflow. Unlike BT, which requires physical positioning of a belt, or NT, which requires calibration of a navigator beam, the radar system is integrated and automated, allowing for a "set it and forget it" approach.
Quantitative analysis showed that for T2-weighted imaging (T2WI), the FT technique was significantly faster than the NT method, with average scan times of approximately 143 seconds versus 179 seconds ($p < 0.01$).
Image Quality and Lesion Conspicuity
While NT remains a strong performer for T2WI sharpness due to direct diaphragm tracking, the FT technique demonstrated superior performance in Diffusion-Weighted Imaging (DWI), particularly at low b-values ($b=50$). In patients, FT-DWI achieved a significantly higher SNR ($705.13 \pm 434.80$) and CNR ($504.41 \pm 400.69$) compared to both BT and NT.
Importantly, ADC values remained consistent across all three techniques, ensuring that the radar method does not interfere with quantitative tissue characterization used to distinguish benign from malignant lesions.
Conclusion
The preliminary results suggest that the contactless radar-trigger technique is a viable and efficient alternative for liver MRI. It offers comparable image quality to traditional methods while improving patient comfort and reducing scan times. Future research with larger cohorts and higher-field (3.0T) systems is needed to further validate these findings across a broader range of pathologies and patient populations.