Flat panel detector-based cone beam CT for dynamic imaging: system evaluation

Ruola Ning, David L. Conover, Yong Yu, Yan Zhang, Weixing Cai, Dong Yang, Xianghua LuProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2006 · DOI

Abstract

The purpose of this study is to characterize a newly built flat panel detector (FPD)-based cone beam CT (CBCT) prototype for dynamic imaging. A CBCT prototype has been designed and constructed by completely modifying a GE HiSpeed Advantage (HSA) CT gantry, incorporating a newly acquired large size real-time FPD (Varian PaxScan 4030CB), a new x-ray generator and a dual focal spot angiography x-ray tube that allows the full coverage of the detector. During data acquisition, the x-ray tube and the FPD can be rotated on the gantry over Nx360 degrees due to integrated slip ring technology with the rotation speed of one second/revolution. With a single scan time of up to 40 seconds , multiple sets of reconstructions can be performed for dynamic studies. The upgrade of this system has been completed. The prototype was used for a series of preliminary phantom studies: different sizes of breast phantoms, a Humanoid chest phantom and scatter correction studies. The results of the phantom studies demonstrate that good image quality can be achieved with this newly built prototype.
BibTeX
@inproceedings{ning2006flat,
  author = {Ruola Ning and David L. Conover and Yong Yu and Yan Zhang and Weixing Cai and Dong Yang and Xianghua Lu},
  title = {{Flat panel detector-based cone beam CT for dynamic imaging: system evaluation}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2006},
  doi = {10.1117/12.655717},
}

Loading related content...

Figure