Flame atomic absorption determination of lead in biological materials using a flow injection on-line separation and preconcentration system based on ion-pair sorbent extraction
Abstract
A method was developed for the trace determination of lead by flame atomic absorption spectrometry with separation and pre-concentration by flow injection on-line ion-pair sorbent extraction. Lead was conveted into its iodoplumbate complex anion, thereby allowing it to be adsorbed on-line on an incorporated C-18 microcolumn as its ion-pair with tetrabutylammonium cation, and subsequently to be eluted by ethanol directly into the nebulizer burner system of the spectrometer. At a consumption of 7-mL sample volume, the procedure resulted in an enhancement factor of 36 and yielded a detection limit of 3 mu g/L (3 sigma) at a sampling frequency of 48/h. The precision (relative Standard deviation, n = 11) was 2.0% at the 300-mu g/L level and 3.1% at the 50-mu g/L, level. The method demonstrated high tolerance to interferences and results obtained for standard reference biological materials were in good agreement with the certified values.