First photoelectron timing error evaluation of a new scintillation detector model
Lists
Only you see this
Abstract
A general timing system model for a scintillation detector developed by Hero et al. (1990) was experimentally evaluated. The detector consists of a scintillator and a photodetector such as a photomultiplier tube (PMT) or an avalanche photodiode. The model uses a Poisson point process to characterize the light output from the scintillator. This timing model was used to simulate a bismuth germanate (BGO) scintillator with a Burle 8575 PMT using first photoelectron timing detection. Evaluation of the model consisted of comparing the root-mean-square error from the simulations with the error from the actual detector system. It is found that the general model compares well with the actual error results for the BGO/8575 PMT detector. >
BibTeX
@article{petrick1991first,
author = {Nicholas Petrick and N.H. Clinthorne and W.L. Rogers and Alfred O. Hero},
title = {{First photoelectron timing error evaluation of a new scintillation detector model}},
journal = {IEEE Transactions on Nuclear Science},
year = {1991},
doi = {10.1109/23.289292},
}Loading related content...