First Laboratory Measurements of a Super-Resolved Compressive Instrument in the Medium Infrared

Donatella Guzzi, Tiziano Bianchi, Marco Corti, Sara Francés González, Cinzia Lastri, Enrico Magli, Vanni Nardino, Christophe Pache, Lorenzo Palombi, Diego Valsesia, Valentina Raimondi2025

TL;DR

This paper presents the first laboratory measurements of a novel super-resolved compressive sensing instrument designed for the medium infrared (MIR) spectrum. By adapting a commercial Digital Micromirror Device (DMD) and employing compressive sensing techniques, the demonstrator achieves significant spatial resolution enhancement (up to 32x) for hot targets, paving the way for more detailed remote sensing applications.

What Problem Does This Paper Address?

Traditional imaging systems often face a trade-off between spatial resolution and the number of spectral bands or the overall data acquisition time. Achieving high spatial resolution across a broad spectral range, especially in the medium infrared (MIR) where thermal emissions are significant, can be challenging and computationally intensive. The authors tackle this by exploring the application of Compressive Sensing (CS) to develop an instrument that can acquire super-resolved images with fewer samples than conventional methods, specifically targeting the MIR spectrum for applications like earth observation.

Methods

The core of the demonstrator is a modified commercial Digital Micromirror Device (DMD). The standard front window of the DMD was replaced with one transparent across the visible to MIR spectrum. This modified DMD acts as a Spatial Light Modulator (SLM) in a whiskbroom spectral imager. The instrument captures images by scanning the scene and projecting it onto the DMD, where a binary pattern is applied. The light is then spectrally split into 10 VIS-NIR bands and 2 MIR bands. After spectral splitting, dichroic mirrors and condensers are used for spatial integration before the signal is filtered and acquired by detectors. The key innovation lies in using the DMD for CS-based super-resolved acquisitions, achieving resolution enhancement factors of up to 32. For image reconstruction, two methods were employed: the classical Total Variation (TV) method and the ISTA-Net, a machine learning-based approach.

Datasets

The paper does not explicitly mention training datasets for machine learning models. The evaluation was performed using laboratory measurements of specifically designed targets. These targets consisted of hot sources (IR lamps) with an emissivity of 0.75 and a maximum temperature of 843 °C. The MIR target used for measurements contained six distinct MIR sources within a single macro-pixel area. This setup served as a benchmark for evaluating the demonstrator's performance in reconstructing super-resolved hot targets.

Key Insights and Ideas

  • DMD as a versatile spectral imager: The modification of a standard DMD to be transparent in the MIR spectrum is a crucial step, enabling its use in a broader range of applications than originally intended.
  • CS for MIR super-resolution: The successful demonstration of CS for achieving super-resolution in the MIR band is significant, as it offers a path to higher detail imaging in this important spectral region.
  • Denoising effect: Both the TV and ISTA-Net reconstruction methods exhibited a denoising effect, producing "cleaner" images than the ground truth, particularly at lower signal-to-noise ratios (SNR) in the MIR channel.
  • ISTA-Net vs. TV: While ISTA-Net showed promise, the classical TV method generally provided better reconstruction quality for the MIR targets in this study.

Key Results

The demonstrator successfully achieved super-resolved acquisitions with resolution enhancement factors of up to 32x in the MIR spectral range. Measurements performed on hot sources demonstrated that the CS approach is effective in reconstructing these targets with enhanced spatial detail. The Signal-to-Noise Ratio (SNR) was evaluated for different super-resolution (SR) values, showing a decrease in SNR as the SR factor increased. For instance, Channel #1 (3.3 µm) had an SNR of 1.4 at SR 16x16 and less than 1 at SR 32x32, while Channel #2 (4.0 µm) maintained an SNR of 14 at SR 16x16 and 2.5 at SR 32x32. The reconstruction quality was generally good, with a noticeable denoising effect, especially for the low-SNR MIR channel.

Limitations and Future Work

The paper acknowledges that the performance of the ISTA-Net method was generally worse than the TV method for MIR reconstruction in this study. The authors also note that the DMD models from Texas Instruments are not inherently compatible with the required large spectral range, necessitating the modification of the front window. Future work could focus on optimizing the ISTA-Net for MIR reconstruction, further improving the SNR performance at higher super-resolution factors, and exploring the application of this technology to real-world earth observation scenarios.

Summary updated

Abstract

In the framework of the SURPRISE EU project, the Compressive Sensing paradigm was applied for the development of a laboratory demonstrator with improved spatial sampling operating from visible up to Medium InfraRed (MIR). The demonstrator, which utilizes a commercial Digital Micromirror Device modified by replacing its front window with one transparent up to MIR, has 10 bands in the VIS-NIR range and 2 bands in the MIR range, showing a super resolution factor up to 32. Measurements performed in the MIR spectral range using hot sources as targets show that CS is effective in reconstructing super-resolved hot targets.