Abstract
In order to obtain stable and high-quality synchrotron radiation photon, the magnets in the storage ring of High Energy Photon Source (HEPS) need to have a stable support and precise positioning. The vibrating wire technique can be used to pre-align the quadrupoles and sextupoles on one girder with high precision to meet the extremely low emittance requirement of HEPS. This thesis introduces the measurement of magnetic center of sextupole using vibrating wire. According to the measurement results, the magnetic field distribution is consistent with theoretical expectation. And vibrating wire has achieved the purpose of measuring the magnetic center and has reached a certain precision.
Introduction
In high-performance particle accelerators like the High Energy Photon Source (HEPS) in China, "emittance"—a measure of the average spread of particle coordinates and momenta—must be kept extremely low to produce high-quality synchrotron radiation. To achieve this, the magnets (quadrupoles and sextupoles) that focus and steer the electron beam must be aligned with incredible accuracy on their support girders.
The primary challenge is that a magnet's mechanical center often deviates from its magnetic center. Traditional mechanical alignment is insufficient; researchers require a method to locate the exact point of zero (or extreme) magnetic field while magnets are mounted. Sextupoles present a particular difficulty due to their complex magnetic field distribution, making high-precision center-finding a significant engineering hurdle.
Experimental Methodology
The researchers utilized the vibrating wire technique to map the magnetic field. The setup involves stretching a thin wire through the magnet's aperture, supported by fixed and free-end benches.
The process follows these steps:
- Lorentz Force Interaction: An alternating current (AC) is passed through the wire. Interaction with the magnet's field causes the wire to vibrate.
- Field Mapping: By sweeping the wire across the aperture and measuring vibration amplitude, the system maps the magnetic induction intensity. For a sextupole, the magnetic center corresponds to the extreme point of a parabolic distribution.
- Systematic Corrections: The team implemented background correction to remove "stray" fields from adjacent magnets and sag correction to account for the ~180μm gravitational droop of the wire.
Data Analysis and Key Insights
A critical finding was the non-trivial nature of background magnetic fields. In multi-magnet girder setups, fields from nearby quadrupoles can shift the perceived horizontal magnetic center by up to 280μm.
As shown in Figure 2, the background field (green triangles) significantly influences the total measured field (red squares). By subtracting this background, the true sextupole field (blue circles) is isolated, allowing for accurate center identification. Additionally, the study confirmed that the magnetic center remains stable across varying excitation currents (80A to 160A), proving the axis is a fixed geometric property.
Results and Performance
The experimental results demonstrated high precision and reliability:
- Repeatability: The system achieved a standard deviation of ±3μm for horizontal measurements and ±4-5μm for vertical measurements.
- Consistency: Data from independent sensors matched within 3μm horizontally and 6μm vertically.
- Practical Alignment: In adjustment trials, the team successfully positioned the sextupole to within 10μm of the target magnetic center, meeting HEPS operational requirements.
Conclusion
The vibrating wire technique has proven effective for locating the magnetic centers of sextupoles with micron-level precision. While vertical measurements show higher sensitivity to environmental noise, the current system meets the stringent alignment needs for HEPS. Future work will focus on measuring quadrupole magnets once technical conditions are restored and improving environmental isolation to further enhance vertical precision.