Feasibility of Golden Angle Spiral Real-Time Phase Contrast MRI at 0.55T: A Single-Center Prospective Study

Salman Pervaiz, Chong Chen, Yingmin Liu, Katherine Binzel, Kelvin Chow, R. Badlishah Ahmad, Yuchi Han, Ning Jin, Juliet VargheseBioengineering 2026

TL;DR

This study demonstrates the feasibility of a new free-breathing, real-time MRI technique for measuring blood flow on a lower-cost, mid-field (0.55T) scanner. This approach shows promising agreement with standard methods, potentially increasing access to advanced cardiac imaging for patients who struggle with breath-holding.

What Problem Does This Paper Address?

Cardiac Magnetic Resonance (CMR) imaging is a powerful tool for diagnosing heart conditions, but higher-field strength scanners (1.5T and 3T) are expensive and less accessible, especially in resource-limited settings. While mid-field (0.55T) scanners offer a more affordable alternative, achieving diagnostic-quality imaging, particularly for dynamic processes like blood flow, remains a challenge. Traditional phase-contrast MRI (PCMR) for flow assessment requires patients to hold their breath, which is difficult for many, including children, the elderly, and those with respiratory or cardiac issues. This can lead to incomplete or inaccurate data. Real-time PCMR (RT-PCMR) overcomes this by acquiring data within a single heartbeat, eliminating the need for breath-holding and capturing beat-to-beat flow variations. However, implementing RT-PCMR on mid-field systems with less powerful gradients has been technically challenging. This paper aims to bridge this gap by developing and validating an RT-PCMR sequence on a commercial 0.55T system.

Methods

The researchers developed a novel RT-PCMR sequence utilizing a dual-density golden-angle spiral readout. This trajectory is designed to efficiently sample the necessary k-space data within a single heartbeat. To further enhance efficiency and image quality on the 0.55T system, they employed SENSE-based compressed sensing (CS) reconstruction. This allows for the acquisition of undersampled data, which is then computationally reconstructed to produce diagnostic images. The reconstruction was performed "inline" using a research prototype pipeline (FIRE framework) on an external GPU-equipped server, allowing for relatively quick processing. For comparison, they used standard breath-held (BH) segmented Cartesian PCMR as the reference standard.

Datasets

The study included data from 33 adult participants: 17 healthy volunteers and 16 patients with cardiovascular disease. A subset of six patients also underwent scanning at both the 0.55T system and a 3T system for cross-field strength comparison. No new benchmark datasets were introduced; the study focused on validating the new sequence against existing, established methods.

Key Insights and Ideas

  • Mid-field MRI is becoming increasingly capable: This study pushes the boundaries of what's possible with lower-field strength MRI, showing that advanced techniques like RT-PCMR can be successfully implemented.
  • Spiral trajectories shine at lower fields: The paper highlights how spiral readouts, which are more efficient at covering k-space, are particularly well-suited for mid-field systems where gradient performance is a limiting factor.
  • Free-breathing flow assessment is within reach: The development of this RT-PCMR sequence offers a viable alternative for patients who cannot reliably perform breath-holds, potentially democratizing access to detailed cardiac flow analysis.
  • The "aha moment" for patients: The ability to capture beat-to-beat flow variations is crucial for understanding conditions like arrhythmias, where traditional methods might miss critical dynamic information. This technique offers a way to see these variations even in patients who struggle with breath-holding.

Key Results

  • Good agreement for key flow parameters: In healthy volunteers, the RT-PCMR sequence showed excellent agreement with BH-PCMR for net flow (ICC > 0.90) and peak flow rate (ICC > 0.90).
  • Slight underestimation of peak velocity: RT-PCMR consistently showed a slight underestimation of peak velocity compared to BH-PCMR, by approximately 7.9% in the aorta and 8.6% in the main pulmonary artery (MPA) in volunteers, and slightly more in patients (up to 12.7% in the aorta). However, these differences were generally within clinically acceptable limits.
  • Comparable results in patients: Similar to volunteers, RT-PCMR showed good agreement for net flow and peak flow rate in patients, with a slight underestimation of peak velocity.
  • Cross-field strength agreement: A small subset of patients scanned at both 0.55T and 3T showed close agreement for most flow parameters, suggesting the technique is robust across different field strengths.
  • Image sharpness: While RT-PCMR images were slightly blurrier than BH-PCMR due to lower spatial resolution, the sharpness was deemed adequate for reliable analysis.

Limitations and Future Work

The authors acknowledge several limitations:

  • Small sample size: The study was conducted on a relatively small cohort, limiting the statistical power and generalizability of the findings.
  • Adult population only: The technique needs to be evaluated in pediatric populations, where higher heart rates and smaller body sizes present different challenges.
  • Absence of flow phantom validation: While validated against a segmented acquisition, direct phantom validation would provide further objective assessment.
  • Limited arrhythmia and shunt evaluation: The study did not include patients with significant arrhythmias or intracardiac shunts, which would be valuable for further assessing RT-PCMR's capabilities in complex scenarios.
  • Parameter optimization: Further optimization of parameters, such as flip angle and regularization for CS reconstruction, could potentially improve performance.

Future work could explore:

  • Deep learning reconstruction: Investigating deep learning-based reconstruction methods could further improve image quality and reconstruction speed.
  • Perturbed spiral trajectories: Exploring more advanced spiral trajectory designs might enhance efficiency and accuracy.
  • Clinical validation in specific populations: Further studies are needed to assess RT-PCMR's performance in pediatric patients, those with arrhythmias, and other challenging patient groups.

Summary updated

Abstract

<b>Background</b>: Real-time phase-contrast magnetic resonance (RT-PCMR) imaging allows free-breathing assessment of blood flow across cardiac valves and vessels. However, the feasibility of free-breathing RT-PCMR on a mid-field (0.55T) MRI system has yet to be established. <b>Aim:</b> The primary objective of this study was to implement a RT-PCMR sequence using a dual-density golden-angle spiral readout with SENSE-based compressed sensing (CS) reconstruction on a 0.55T MRI system. The secondary objective was to evaluate the feasibility of this approach in an adult cohort comprising healthy volunteers and patients with cardiovascular disease. <b>Materials and Methods:</b> Data from 33 participants were included in the flow quantification analysis (healthy volunteers: <i>n</i> = 17, 9 females, mean age 30.4 ± 14.6 years; patients: <i>n</i> = 16, 11 females, mean age 45.9 ± 17.4 years), with breath-held (BH) segmented Cartesian PCMR used as the reference standard. <b>Results:</b> In volunteers, RT-PCMR showed good agreement for net flow, peak flow rate, and pulmonary-systemic flow ratio (Qp/Qs), without significant bias (<i>p</i> > 0.05) and slightly underestimated peak velocity [7.9% in the aorta and 8.6% in the main pulmonary artery (MPA)]. In patients, RT-PCMR slightly underestimated peak flow rate (aorta, 6.2%; MPA; 4.6%) and peak velocity (aorta,12.7%; MPA, 10.4%). A sub-analysis of six patients scanned at both 0.55T and 3T showed close agreement between field strengths. <b>Conclusions:</b> These results demonstrate the feasibility of our RT-PCMR sequence on a commercial 0.55T system.
BibTeX
@inproceedings{pervaiz2026feasibility,
  author = {Salman Pervaiz and Chong Chen and Yingmin Liu and Katherine Binzel and Kelvin Chow and R. Badlishah Ahmad and Yuchi Han and Rizwan Ahmad and Ning Jin and Juliet Varghese},
  title = {{Feasibility of Golden Angle Spiral Real-Time Phase Contrast MRI at 0.55T: A Single-Center Prospective Study}},
  booktitle = {Bioengineering},
  year = {2026},
  doi = {10.3390/bioengineering13020166},
}
Figure