Fabrication of Stent-Like Mesh Structures Using Synchronized Scan Rotation Lithography and Wet Etching

Toshiyuki Horiuchi, Kaiki Ito, Yuta Suzuki, Akira Yanagida, Hiroshi KobayashiJournal of Photopolymer Science and Technology 2020 · DOI

TL;DR

Researchers have developed a novel method for fabricating intricate stent-like mesh structures from stainless steel pipes using synchronized scan rotation lithography and wet etching. This technique offers a promising pathway for creating customized, high-precision medical implants with potential for improved patient outcomes.

What Problem Does This Paper Address?

Degradation and damage to blood vessels are serious health concerns, and medical stents play a crucial role in supporting and repairing these vital structures. While various stent designs and fabrication methods exist, there's a continuous need for more precise, cost-effective, and adaptable manufacturing techniques to produce trustworthy and inexpensive stents. This paper addresses the challenge of creating complex, micro-scale cylindrical mesh structures with high accuracy and homogeneity, which is essential for effective stent function.

Methods

The core of this work lies in a two-step fabrication process:

  1. Synchronized Scan Rotation Lithography: This novel lithography technique involves coating a thin-walled stainless steel pipe with a photosensitive resist. A flat reticle containing the desired stent pattern is then projected onto the pipe. The key innovation is the synchronized linear scanning of the reticle and the rotation of the pipe. An oblong slit on the reticle limits the exposure area to the pipe's ridge, ensuring precise pattern transfer across the entire cylindrical surface without gaps or overlaps. A negative resist was chosen for its robustness against double exposure during the stitching process.
  2. Wet Chemical Etching: Following lithography and resist development, the patterned pipe is immersed in an aqueous solution of ferric chloride. The resist patterns act as an etching mask, protecting the underlying material. The exposed areas are selectively etched away, gradually forming the stent-like mesh structure. The etching process is carefully controlled in terms of temperature, stirring, and time to achieve the desired mesh dimensions.

Datasets

Not discussed in the paper. The study focuses on the fabrication process itself, using a stainless steel pipe as the substrate and a reticle with stent patterns as the design input.

Key Insights and Ideas

  • Precise Cylindrical Patterning: The synchronized scan rotation lithography effectively overcomes the challenges of patterning complex shapes onto curved surfaces, achieving a continuous and accurate transfer of intricate mesh designs.
  • Undercutting as a Design Factor: The wet etching process inherently involves undercutting the resist mask. While this leads to a reduction in mesh width compared to the resist pattern, understanding and controlling this phenomenon is crucial for achieving the target dimensions.
  • Negative Resist Advantage: The use of a negative resist proved beneficial for handling the overlapping edges during the synchronized scanning, as it showed less deformation compared to positive resists under double exposure.
  • Correlation Between Resist and Mesh Width: A strong positive correlation was observed between the width of the resist patterns and the final etched mesh widths, indicating that lithography accuracy directly influences the final stent dimensions.

Key Results

  • A stent-like mesh structure with a mean mesh width of 108 ± 12.4 µm (3σ) was successfully fabricated from a SUS 304 stainless steel pipe.
  • The fabrication process demonstrated good homogeneity, with fluctuations in mesh widths within 3σ deviations of 11.4 µm (axial) and 12.4 µm (circumferential).
  • The width of the etched mesh was consistently narrower than the resist pattern width due to the undercutting effect during wet etching. The mean difference was found to be 78 µm.
  • The study provided a detailed analysis of the cross-sectional profile of the undercut, estimating the width at the outer surface (ws) to be 60.2 µm and the inclined part width (wi) to be 24.4 µm.

Limitations and Future Work

  • Fluctuation Reduction: While the obtained homogeneity is considered decent for a developing stage, the authors acknowledge that fluctuations in mesh widths need to be further reduced for higher accuracy.
  • Patterning Accuracy Improvement: The study highlights the necessity of improving the patterning accuracy of the lithography process as a primary step towards achieving more precise structures.
  • Etching Homogeneity: Further improvements in the etching process are also suggested to enhance overall width homogeneity.
  • Future Work: The authors intend to continue refining the lithography and etching processes to achieve even greater precision and homogeneity in the fabricated stent-like structures.

Summary updated

Abstract

Stent-like complicated cylindrical structures were made with pipes of stainless-steel SUS 304 using new lithography and wet chemical etching. In the new lithography, patterns on a flat reticle were printed on a pipe coated with a resist film by synchronously scanning the reticle linearly and rotating the pipe around the axis, and limiting the momentary exposure area on the top ridge of the pipe by placing an oblong slit on the reticle in parallel to the pipe axis. The patterned pipe was wetly etched in an aqueous solution of ferric chloride using the resist patterns as etching masks. Because the etching was progressed equally in all the directions from the resist pattern edges, masked parts under the resist patterns were also gradually etched from the pattern edges and undercut during the pipe was penetrated through the wall. Caused by the undercut, obtained mesh widths became narrower than the resist pattern widths. However, a stent-like mesh structure with widths of 108±12.4 (3σ) µm was decently fabricated by appropriately controlling the resist pattern width and the etching time. To attain higher accuracy in the future, cross section profiles of the mesh and relationship between resist pattern widths and mesh widths were discussed in detail.
BibTeX
@article{horiuchi2020fabrication,
  author = {Toshiyuki Horiuchi and Kaiki Ito and Yuta Suzuki and Akira Yanagida and Hiroshi Kobayashi},
  title = {{Fabrication of Stent-Like Mesh Structures Using Synchronized Scan Rotation Lithography and Wet Etching}},
  journal = {Journal of Photopolymer Science and Technology},
  year = {2020},
  doi = {10.2494/photopolymer.33.361},
}

Loading related content...