Introduction
The management of ship ballast water is critical for preventing the introduction of invasive aquatic species into non-native ecosystems. Common treatment methods, such as electrolysis and chlorination, produce Total Residual Oxidants (TRO) that must be monitored to ensure effective disinfection while preventing environmental damage upon discharge. While the DPD colorimetric method is the current industry standard, it suffers from reagent dependency and sensitivity to water turbidity. Electrochemical sensors offer a faster, reagent-free alternative, but their reliability is often compromised by the instability of traditional liquid-junction reference electrodes in high-salinity marine environments. This study explores the development of a gel-type internal electrolyte using hydroxyethyl-cellulose (HEC) to enhance the durability and performance of TRO sensors.
Materials and Methods
The research focused on replacing standard 3.5 M KCl liquid electrolytes with a polymer gel to mitigate leakage and contamination. Gel electrolytes were synthesized by combining 3.5 M KCl with glycerol and varying concentrations of HEC (ranging from 5 to 15 wt%).
The experimental setup utilized a three-electrode system integrated into a TRO sensor. This system comprised a Gold (Au) working electrode, a Platinum (Pt) counter electrode, and the newly developed Ag/AgCl reference electrode containing the HEC gel. Performance was evaluated using amperometry at a constant potential of 50 mV.
Results and Discussion
Electrolyte Characterization and Durability
Initial testing focused on the trade-off between polymer content and ionic conductivity. While higher HEC concentrations improved the physical stability of the gel, they simultaneously reduced ionic conductivity. A concentration of 10% HEC was identified as the optimal "sweet spot," providing sufficient conductivity (192.4 mS/cm) while significantly reducing KCl leakage compared to lower HEC formulations during a 60-day immersion test in deionized water.
Sensor Performance and Salinity Effects
The sensor demonstrated a rapid response time, reaching a stable current within 5–10 seconds across various TRO concentrations (0–15 mg/L). Crucially, the sensor remained functional across a wide range of salinities, from 0.2 PSU (fresh water) to 30 PSU (seawater).
As shown in the performance data, the sensor exhibits a linear relationship between current density and TRO concentration. However, the sensitivity (slope of the response) increases with salinity. This phenomenon is attributed to the reduced electrical resistance of the bulk solution at higher salt concentrations, necessitating the use of salinity compensation algorithms for accurate field measurements.
Conclusion
The development of an HEC-based gel electrolyte significantly improves the robustness of electrochemical TRO sensors for marine applications. By optimizing the polymer content to 10%, researchers achieved a balance of high ionic conductivity and long-term stability. Future work will focus on integrating salinity compensation software and addressing real-world challenges such as biofouling and mineral scaling on electrode surfaces.