Fabrication of Flexible Microneedle Array Electrodes for Wearable Bio-Signal Recording

Lei Ren, Shujia Xu, Jie Gao, Zi Lin, Zhipeng Chen, Bin Liu, Liang Liang, Lelun Jiang2018 · DOI

TL;DR

Researchers have developed a novel, flexible microneedle array electrode (MAE) for wearable bio-signal monitoring. This MAE offers improved skin contact and reduced impedance compared to traditional electrodes, enabling high-fidelity recording of ECG, EMG, and EEG signals with minimal artifacts, even after repeated bending.

What Problem Does This Paper Address?

The increasing demand for home health care necessitates reliable and comfortable bio-signal monitoring. Traditional Ag/AgCl electrodes, while widely used, require skin preparation and gel, which can cause irritation, dry out, and degrade signal quality over time. Existing flexible electrodes often struggle to maintain consistent contact with curved skin surfaces, leading to motion artifacts. Microneedle array electrodes (MAEs) offer a promising solution by piercing the stratum corneum to establish a stable interface, but many are fabricated on rigid substrates, limiting their conformability. This paper addresses the need for a flexible, comfortable, and high-performance MAE suitable for long-term wearable bio-signal monitoring.

Methods

The researchers employed a two-pronged fabrication approach:

  • Laser-Direct Writing (LDW): This technique was used to create conductive patterns on a flexible polyethylene terephthalate (PET) substrate. Silver ink was applied and then sintered by a focused laser beam, allowing for precise, maskless, and low-cost pattern generation. The serpentine design of the conductive pattern was chosen to enhance mechanical deformability and prevent fracture.
  • Magneto-Rheological Drawing Lithography (MRDL): This method was used for the one-step fabrication of the microneedle array directly onto the conductive pattern. Curable magneto-rheological fluid (CMRF) was drawn into microneedle shapes using an external magnetic field and a pillar array, then solidified. This process is described as rapid and efficient.

Finally, the microneedle array surface was sputter-coated with Ti/Au films to enhance conductivity and biocompatibility.

Datasets

Not discussed in the paper. The study used human volunteers for bio-signal recording.

Key Insights and Ideas

  • Synergistic Fabrication: The combination of LDW for conductive patterns and MRDL for microneedles offers a streamlined and effective method for creating flexible MAEs.
  • Conformability is Key: The inherent flexibility of the PET substrate, coupled with the microneedle design, allows the MAE to conform closely to curved skin surfaces, significantly improving contact and reducing motion artifacts.
  • Reduced Skin Impedance: The microneedles effectively bypass the high-impedance stratum corneum, leading to a lower electrode-skin interface impedance (EII) compared to traditional and other flexible dry electrodes.
  • Robustness to Bending: The MAE maintains low EII even after 50 cycles of bending, indicating good mechanical durability for wearable applications.
  • High-Fidelity Signal Recording: The MAE demonstrated superior performance in recording ECG, EMG, and EEG signals, particularly in dynamic conditions, with clearer waveforms and less noise than Ag/AgCl and FDEs.

Key Results

  • Lowest EII: The flexible MAE exhibited the lowest electrode-skin interface impedance, even after 50 bending cycles, compared to Ag/AgCl and flexible dry electrodes (FDE).
  • Improved ECG Recording: Dynamic ECG signals recorded by the flexible MAE were more distinguishable and had less noise than those from Ag/AgCl and FDE.
  • Suitable for Small Muscles: The flexible MAE's smaller contact area (approx. 100 mm²) compared to Ag/AgCl electrodes (approx. 176 mm²) makes it more suitable for recording EMG from small muscles.
  • Microneedle Dimensions: Microneedles had an average height of 500 ± 10 µm, tip radius of 10 ± 5 µm, and base diameter of 350 ± 15 µm, optimized for skin penetration and bonding.

Limitations and Future Work

  • Limited Volunteer Pool: The study involved only three healthy volunteers, limiting the generalizability of the findings.
  • Long-Term Wearability: While the MAE shows promise for long-term monitoring due to its comfort and low impedance, extensive studies on long-term wearability and potential skin reactions over extended periods would be beneficial.
  • Customization: The paper mentions that microneedle distribution and number can be customized, suggesting future work could explore optimized designs for specific bio-signals or anatomical locations.
  • Further Signal Analysis: While ECG, EMG, and EEG were tested, exploring other bio-signals and more complex physiological conditions could expand the MAE's application scope.