Extending the Lifetime of MEDA Biochips by Selective Sensing on Microelectrodes
Abstract
A digital microfluidic biochip (DMFB) enables miniaturization of immunoassays, point-of-care clinical diagnostics, and DNA sequencing. A recent generation of DMFBs uses a micro-electrode-dot-array (MEDA) architecture, which provides fine-grained control of droplets and real-time droplet sensing using the CMOS technology. However, microelectrodes in a MEDA biochip degrade when they are charged and discharged frequently during bioassay execution. In this article, we first make the key observation that the droplet-sensing operations contribute up to 94% of all microelectrode actuation in MEDA. Consequently, to reduce the number of droplet-sensing operations, we present a new microelectrode cell (MC) design as well as a selective-sensing method such that only a small fraction of microelectrodes perform droplet sensing during bioassay execution. The selection of microelectrodes that need to perform the droplet sensing is based on an analysis of experimental data. A comprehensive set of simulation results show that the total number of droplet-sensing operations is reduced to only 0.7%, which prolongs the lifespan of a MEDA biochip by 11× without any impact on bioassay time-to-response.
BibTeX
@article{liang2020extending,
author = {Tung-Che Liang and Zhanwei Zhong and Miroslav Pajic and Krishnendu Chakrabarty},
title = {{Extending the Lifetime of MEDA Biochips by Selective Sensing on Microelectrodes}},
journal = {IEEE Trans. Comput. Aided Des. Integr. Circuits Syst.},
year = {2020},
doi = {10.1109/tcad.2020.3012746},
}Ask Althea to put this paper to the test.