Evaluating the performance of an implantable near infrared spectroscopy sensor for monitoring spinal cord tissue oxygenation

Garrett Frank, Kitty So, Jay Ethridge, Neda Manouchehri, Shi Haida, Katharina Raschdorf, Ali N. Zaidi, Carly Levinsky, Avril Billingsley, J. Ross Mitchell, Dane W. Grasse, Vishnoukumaar Sivaji, Rahul Saini, Femke Streijger, Brian K. Kwon2025

Abstract

We developed an implantable near-infrared spectroscopy sensor (D-NIRS) to monitor spinal cord tissue oxygenation following injury. This study evaluates the D-NIRS sensor on the porcine spinal cord by comparing its measurements with intraparenchymal (IP) oxygenation sensors and arterial blood saturation levels during hypoxia. Two D-NIRS sensors and two IP probes (Oxford Optronix, Adderbury, UK) were implanted in the thoracic spinal cord of 6 Yucatan Minipigs. Hypoxia was induced to achieve a target minimum of 70% arterial oxygen saturation. Carotid blood samples were collected to assess arterial oxygen saturation (SaO2). Change in D-NIRS tissue oxygenation index (TOI) due to hypoxia strongly correlated to change in SaO2 (r=0.848) and change in TOI relative to SaO2 between NIRS devices (r=0.787). IP measurements were more variable (r=0.562 and r=0.453 respectively). The D-NIRS sensors appear to reliably measure spinal cord tissue oxygenation, showing a strong correlation to systemic blood oxygenation.
BibTeX
@misc{frank2025evaluating,
  author = {Garrett Frank and Kitty So and Jay Ethridge and Neda Manouchehri and Shi Haida and Katharina Raschdorf and Ali N. Zaidi and Carly Levinsky and Avril Billingsley and J. Ross Mitchell and Dane W. Grasse and Vishnoukumaar Sivaji and Rahul Saini and Femke Streijger and Brian K. Kwon},
  title = {{Evaluating the performance of an implantable near infrared spectroscopy sensor for monitoring spinal cord tissue oxygenation}},
  year = {2025},
  doi = {10.1117/12.3057158},
}

Loading related content...

Figure