Estimation of tumor coverage after RF ablation of hepatocellular carcinoma using single 2D image slices

Nicole Varble, Ming Li, Laetitia Saccenti, Tabea Borde, Antonio Arrichiello, Anna Christou, Katerina Lee, Lindsey Hazen, Sheng Xu, Riccardo Lencioni, Bradford J. WoodInt. J. Comput. Assist. Radiol. Surg 2025

Introduction

Radiofrequency ablation (RFA) is a primary treatment for hepatocellular carcinoma (HCC), utilizing thermal energy to destroy malignant tissue. The clinical success of RFA is defined by "technical success," which requires the ablation zone to completely cover the target tumor. Currently, clinicians rely on either subjective visual inspections or time-consuming 3D software analysis involving manual segmentation and elastic registration. This study proposes an AI-based method to estimate tumor coverage using single 2D CT image slices, aiming to streamline the assessment process.

Methodology

The researchers developed a "feature subtraction" architecture to bypass the complexities of 3D processing. The model takes two 2D axial CT slices as input: one from the center of the pre-treatment tumor and one from the center of the post-treatment ablation zone (typically 1 month post-procedure).

Features are extracted using a pre-trained VGG16 convolutional neural network, resulting in 25,088 features per image. Instead of a standard Siamese network, the model performs an element-wise subtraction of these feature vectors. This "difference vector" is then processed by a Random Forest classifier to sort cases into four coverage categories (<50%, 50–70%, 70–90%, and ≥90%) or a regression model to predict a specific percentage.

Overview of the feature extraction and predictive model architecture.

Results and Performance

The study utilized data from 182 patients in the OPTIMA trial, expanded via data augmentation to 728 image pairs. The classification model achieved an AUC of 0.86 and an accuracy of 69% on the test cohort. Regression models yielded a Mean Absolute Error (MAE) of 9.8%.

Performance analysis showed that the model was most accurate for coverage levels between 60% and 85%. However, it tended to overestimate coverage for poor results and underestimate it for high-coverage cases.

Confusion matrices illustrating the model's predictive accuracy on original and noise-augmented datasets.

Discussion and Limitations

A critical finding was the model's failure to generalize to an external validation cohort, where performance was no better than random. This "generalization gap" is attributed to variations in CT protocols, tumor volumes, and the timing of follow-up scans (e.g., 21 days vs. 28 days).

Furthermore, the 2D approach inherently lacks spatial context in the cranial-caudal direction. While the "center slice" hypothesis simplifies the workflow, it may miss failures occurring outside the primary axial plane. Future research will focus on incorporating multi-institutional data and exploring "2.5D" models that utilize adjacent slices to capture more spatial information without the full overhead of 3D analysis.

Summary updated

BibTeX
@article{varble2025estimation,
  author = {Nicole Varble and Ming Li and Laetitia Saccenti and Tabea Borde and Antonio Arrichiello and Anna Christou and Katerina Lee and Lindsey Hazen and Sheng Xu and Riccardo Lencioni and Bradford J. Wood},
  title = {{Estimation of tumor coverage after RF ablation of hepatocellular carcinoma using single 2D image slices}},
  journal = {Int. J. Comput. Assist. Radiol. Surg},
  year = {2025},
  doi = {10.1007/s11548-025-03423-z},
}
Figure