TL;DR
This paper introduces a novel, non-contact method for estimating respiratory rate (RR) using thermal imaging. By leveraging deep learning for facial detection and a new "Respiratory Quality Index" (RQI), the system can accurately track breathing from thermal facial data, offering a promising solution for emergency monitoring where traditional methods are impractical.
What Problem Does This Paper Address?
Accurate and continuous monitoring of respiratory rate (RR) is crucial for assessing a patient's health status, especially in emergency situations. Traditional methods often involve attaching sensors to the body, which can be uncomfortable, lead to measurement errors due to movement, and are unsuitable for patients with severe trauma. While non-contact methods are being developed, many still rely on specific facial features like the nostrils or require subjects to remain perfectly still, limiting their applicability in dynamic environments like ambulances. This paper aims to bridge this gap by developing a robust, non-contact RR estimation system that only requires a thermal image of the entire face, making it more versatile for real-world emergency scenarios.
Methods
The proposed system employs a multi-stage approach:
- Facial Region Detection: The system first utilizes YOLOv3, a state-of-the-art deep learning object detection model, to identify and isolate the facial region within thermal images. This is a key advantage as it doesn't require precise localization of nostrils or mouth.
- Subregion Signal Extraction: The detected facial region is then divided into a grid (4x6 subregions). For each subregion, the system extracts signal intensity over time, representing temperature fluctuations.
- Signal Processing: A rectangular window is applied to the signal for frequency resolution, and the signal is detrended to remove linear trends, improving accuracy in the low-frequency respiratory range.
- Respiratory Likelihood Index (RQI): This is the core novelty. The RQI quantifies how likely a given subregion is to contain respiratory-related signals. It's calculated based on two components:
- Spectrum Index (SI): Analyzes the power spectrum density of the signal. Respiratory signals typically show a prominent peak in the breath-related frequency (BF) band (0.1-2 Hz) with less activity in the high-frequency (HF) band. SI captures this characteristic.
- Agreement between Time and Frequency Domain RR Estimates: The RQI also incorporates a term that measures the agreement between RR estimates derived from time-domain (zero-crossing) and frequency-domain (power spectrum) analyses. Higher agreement suggests a more reliable respiratory signal.
- A sigmoid function is used to map these components into a [0, 1] range, where 1 indicates a highly likely respiratory signal.
- RR Estimation: The subregion with the maximum RQI is automatically selected as the respiratory-related region, and its estimated RR (derived from frequency domain analysis) is taken as the subject's respiratory rate.
Datasets
- Training Data: 772 thermal images captured by a Boson 320 camera were used for YOLOv3 training. These images included 12 subjects (20-24 years old) in both seated and supine positions.
- Evaluation Data: 77 thermal images from 7 healthy subjects (4 female, 3 male) were used for face detection validation. For RR estimation, four female and three male subjects (20-24 years old) participated, with four 15-second measurements each, instructed to set their RR to 15, 20, 25, and 30 bpm.
Key Insights and Ideas
- Facial Region Sufficiency: The study demonstrates that the entire facial region, not just specific points like the nostrils, contains sufficient thermal information to estimate RR. This significantly simplifies the system's requirements.
- Respiratory Quality Index (RQI): The RQI is a clever way to fuse spectral characteristics and temporal consistency to identify the most reliable respiratory signals within the face. It moves beyond simple signal amplitude to assess signal quality.
- Robustness to Non-Ideal Conditions: By focusing on the entire facial region and using a robust detection model like YOLOv3, the system shows promise for handling variations in subject pose and potentially less ideal imaging conditions compared to methods requiring precise feature tracking.
- Potential for Ambulance Use: The non-contact nature and reliance on thermal imaging make this method particularly suitable for environments like ambulances, where subject movement and motion artifacts are common challenges for traditional sensors.
Key Results
- Face Detection: The face detection achieved good agreement with ground truth, with an average IoU of 0.86 and an IoU0.70 of 1.00.
- RR Estimation Accuracy: The proposed method achieved a Mean Absolute Error (MAE) of 0.66 bpm when compared to a piezoelectric respiratory measurement belt.
- Bland-Altman Analysis: The bias average was 0.19 bpm, and the 95% limits of agreement ranged from -1.9 to 2.3 bpm, indicating good agreement with the ground truth.
Limitations and Future Work
- Limited Subject Count: The statistical analysis was based on a relatively small number of subjects (7 for RR estimation).
- Ambulance Shaking: The effects of ambulance shaking on the system's performance were not explicitly investigated.
- Apnea Detection: The current system's estimated RR range is 6-120 bpm, meaning it cannot currently detect apnea (cessation of breathing).
- Future Work: The authors plan to investigate the distribution of RQI to identify apnea and evaluate the method in a clinical study. They also suggest exploring machine learning methods (like SVM) to better utilize the features within RQI for subregion determination and to detect patients and clinicians as separate classes in real-world emergency scenarios.
Summary updated