Enhanced sensitivity of VEGF detection using catalase-mediated chemiluminescence immunoassay based on CdTe QD/H2O2 system

Fahimeh Ghavamipour, Hossein Rahmani, Maryam Shanehsaz, Khosro Khajeh, Manouchehr Mirshahi, Reza H. SajediJournal of Nanobiotechnology 2020 · DOI

Abstract

The novel immunoassay proposes a highly sensitive, specific, and stable method for very low levels detection of VEGF that can be used in the primary diagnosis of tumors. With the well-designed sensing platform, this approach has a broad potential to be applied for quantitative analysis of numerous disease-related protein biomarkers for which antibodies are available.

Introduction

Vascular Endothelial Growth Factor (VEGF) serves as a critical biomarker for angiogenesis, the process by which tumors develop new blood supply networks. Accurate detection of VEGF at ultra-low concentrations is essential for early-stage cancer diagnosis and monitoring therapeutic efficacy. While the Enzyme-Linked Immunosorbent Assay (ELISA) is the clinical standard, conventional colorimetric versions utilizing Horseradish Peroxidase (HRP) often lack the sensitivity required for early detection. This study introduces a novel chemiluminescence immunoassay designed to overcome these limitations by redesigning the signal transduction pathway.

Methodology and Innovation

The researchers implemented three primary innovations to enhance the traditional ELISA framework:

  1. High-Turnover Reporter Enzyme: Catalase (CAT) was selected to replace HRP. Catalase boasts one of the highest known enzymatic turnover numbers, decomposing hydrogen peroxide ($H_2O_2$) approximately 240 times faster than HRP.
  2. Dextran Scaffolding: A Dextran T500kD scaffold was employed as a 3D bridge to link multiple catalase enzymes to a single VEGF antigen. This high-density loading significantly amplifies the signal generated per binding event.
  3. Quantum Dot Signal Transduction: TGA-capped CdTe Quantum Dots (QDs) were used to generate chemiluminescence upon reaction with $H_2O_2$.
Characterization of the CAT-VEGF-Dextran conjugate via SDS-PAGE, size exclusion chromatography, and activity assays.

The resulting CAT-VEGF-Dextran conjugate was characterized to ensure successful assembly and maintained enzymatic activity. As shown in Figure 1, size exclusion chromatography and gel electrophoresis confirmed the formation of the high-molecular-weight complex.

Assay Mechanism

The platform operates as a competitive immunoassay. Sample VEGF competes with the engineered CAT-VEGF-Dextran conjugate for binding sites on an anti-VEGF antibody-coated plate. Following a wash step, $H_2O_2$ is introduced. The bound catalase consumes the $H_2O_2$ substrate. Finally, QDs are added, and the remaining $H_2O_2$ triggers a chemiluminescence signal. Consequently, the light intensity is inversely proportional to the concentration of VEGF in the original sample.

Results and Discussion

The assay demonstrated exceptional performance metrics. The Limit of Detection (LOD) was established at 0.5 pg/mL, representing a ten-fold improvement over standard HRP-based colorimetric ELISAs (5 pg/mL). The system maintained a wide linear response range from 2 to 35,000 pg/mL.

Chemiluminescence (CL) intensity response showing the inverse relationship with CAT-VEGF conjugate concentration (a) and reaction time (b).

As illustrated in Figure 6, the chemiluminescence intensity decreases predictably as the concentration of the catalase-labeled conjugate increases, validating the "inverse logic" of the detection system. Clinical validation using human serum samples showed a near-perfect correlation ($R^2 = 0.997$) with established commercial kits, alongside high recovery rates (97–107%) and excellent precision (RSD as low as 1.2%).

Conclusion

By leveraging the extreme catalytic efficiency of catalase and the unique optical properties of quantum dots, this study provides a robust platform for ultra-sensitive protein detection. While the synthesis of specialized components adds complexity, the modular nature of the assay suggests it could be adapted for a wide array of low-abundance disease biomarkers, offering a powerful tool for early clinical intervention.

Summary updated

Abstract

The novel immunoassay proposes a highly sensitive, specific, and stable method for very low levels detection of VEGF that can be used in the primary diagnosis of tumors. With the well-designed sensing platform, this approach has a broad potential to be applied for quantitative analysis of numerous disease-related protein biomarkers for which antibodies are available.