Effects of Regularisation Priors and Anatomical Partial Volume Correction on Dynamic PET Data
Lists
Only you see this
Abstract
Dynamic PET provides temporal information about the tracer uptake. However, each PET frame has usually low statistics, resulting in noisy images. Furthermore, PET images suffer from partial volume effects. The goal of this study is to understand the effects of prior regularisation on dynamic PET data and subsequent anatomical partial volume correction. The Median Root Prior (MRP) regularisation method was used in this work during reconstruction. The quantification and noise in image-domain and time-domain (time-activity curves) as well as the impact on parametric images is assessed and compared with Ordinary Poisson Ordered Subset Expectation Maximisation (OP-OSEM) reconstruction with and without Gaussian filter. This study shows the improvement in PET images and time-activity curves (TAC) in terms of noise as well as in the parametric images when using prior regularisation in dynamic PET data. Anatomical partial volume correction improves the TAC and consequently, parametric images. Therefore, the use of MRP with anatomical partial volume correction is of interest for dynamic PET studies.