Effects of multi-vendor gradient non-linearity correction on breast tumor ADC measurements in the ACRIN 6698 trial
Abstract
Motivation: MRI scanner gradient non-linearity (GNL) is a known source of variability and spatially-dependent bias in quantitative ADC measurements derived from diffusion-weighted imaging (DWI). Goal(s): Evaluate the effects of GNL correction on DWI from the ACRIN 6698 multi-center trial, which investigated ADC as a marker breast cancer response in patients receiving neoadjuvant therapy. Approach: Retrospective GNL correction was performed on the ACRIN 6698 DWI dataResults: Percent decrease in mean tumor ADC post-GNC ranged from 0.5%-11% for different MRI gradient sets, illustrating that GNL can confer significant bias to ADC measurements and should be corrected in multi-center clinical DWI trials.Impact: Implementation of gradient non-linearity correction to reduce bias and variability of tumor ADC measurements from DWI data acquired in multi-center, multi-platform oncology trials may improve the quantitative utility of ADC for characterizing response to treatment.
BibTeX
@inproceedings{wilmes2024effects,
author = {Lisa J. Wilmes and Wen J. Li and Judith Zimmermann and David C. Newitt and Dariya Malyarenko and Jiachao Liang and Patrick J. Bolan and Savannah C. Partridge and Thomas L. Chenevert and Nola M. Hylton and Nola Hylton},
title = {{Effects of multi-vendor gradient non-linearity correction on breast tumor ADC measurements in the ACRIN 6698 trial}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2024/0206},
}Ask Althea to put this paper to the test.