Effects of angular range on image quality of chest digital tomosynthesis

Haenghwa Lee, Yeseul Kim, Sunghoon Choi, Donghoon Lee, Seungyeon Choi, Hee-Joung KimProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2016

Abstract

Chest digital tomosynthesis (CDT) is a new 3D imaging technique that can be expected to improve clinical diagnosis over conventional chest radiography. We investigated the effect of the angular range of data acquisition on the image quality using newly developed CDT system. The four different acquisition sets were studied using ±15°, ±20°, ±30°, and ±35° angular ranges with 21 projection views (PVs). The point spread function (PSF), modulation transfer function (MTF), artifact spread function (ASF), and normalized contrast-to-noise ratio (CNR) were used to evaluate the image quality. We found that increasing angular ranges improved vertical resolution. The results indicated that there was the opposite relationship of the CNR with angular range for the two tissue types. While CNR for heart tissue increased with increasing angular range, CNR for spine bone decreased. The results showed that the angular range is an important parameter for the CDT exam.