Effectiveness and Reliability Assessment on Head Motion Capturing and Correction (MoCAP)
Abstract
Head motion monitoring and compensation during MRI is essential to imaging quality and success rates of acquisition. MoCAP is a novel real-time head motion monitoring and correction technique utilizing structured light to perform prospective motion correction by adjusting MR gradient. We perform effectiveness and reliability assessment on MoCAP in structural and functional MRI. MoCAP can significantly improve quality of the two modalities, reducing motion artifacts and enhancing validity and reliability of post-processing results. MoCAP is promising in the MRI field for special populations like children and patients with difficulty in keeping still during scan.
BibTeX
@inproceedings{gu2024effectiveness,
author = {Zhuoyang Gu and Lianghu Guo and Qing Yang and Xinyi Cai and Tianli Tao and Sifan He and Hua Jiang and Haifeng Tang and Qian Wang and Xiaopeng Zong and Dinggang Shen and Qiang He and Han Zhang},
title = {{Effectiveness and Reliability Assessment on Head Motion Capturing and Correction (MoCAP)}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},