[Effect on inspiratory cycling-off synchronization with different expiratory trigger sensitivity level during pressure-support ventilation: a lung model study].
Lists
Only you see this
Abstract
There are significant differences in performances and cycling-off in obstructive and restrictive conditions among 2 bilevel ventilators. And it may be due to the software of algorithm. Modifications of ETS can affect the ventilatory parameters to improve patient-ventilator synchrony in the presence of air leaks. The Auto-Trak system during PSV show similar results and better synchronzation compared with conventional adjustments.
BibTeX
@inproceedings{chen2014effect,
author = {Yuqing Chen and Xingyi Zhang and Kewen Cheng and Xin Zhou},
title = {{[Effect on inspiratory cycling-off synchronization with different expiratory trigger sensitivity level during pressure-support ventilation: a lung model study].}},
booktitle = {PubMed},
year = {2014},