TL;DR
This paper presents a novel system for denoising electrocardiogram (ECG) signals recorded from the ear canal (ear-ECG). By combining in-ear heart sound recordings for precise cardiac cycle timing with an Extended Kalman Filter and a personalized ECG dynamical model, they achieve significant noise reduction and accurate ECG signal recovery, paving the way for unobtrusive, continuous heart monitoring.
What Problem Does This Paper Address?
The increasing integration of wearable technology, particularly smartwatches and in-ear devices (earables), offers exciting possibilities for continuous health monitoring. Earables, due to their stable placement and user familiarity, are particularly attractive for recording electrocardiogram (ECG) signals. However, recording ECGs from the ear presents a significant challenge: the signal-to-noise ratio (SNR) is extremely low. This is because other biosignals, such as brain (EEG) and muscle (EMG) activity, have much higher amplitudes and overlap with the ECG's frequency spectrum at this location. Existing denoising methods, often relying on black-box autoencoders, struggle to recover accurate ECG morphology and lack explainability. This paper aims to address this by developing a more robust, explainable, and accurate ear-ECG denoising system.
Methods
The core of this work lies in a multi-modal denoising approach.
- In-Ear Heart Sound Recording: The system utilizes in-ear microphones to record heart sounds. Specifically, the S1 heart sound (corresponding to the QRS complex) is used to precisely timestamp individual cardiac cycles. This is achieved by leveraging bone conduction and the occlusion effect within the ear canal.
- ECG Dynamical Model Fitting: Initially, the recorded ear-ECG signals are averaged to create a single, ensemble-average cardiac cycle. This averaged cycle is then used to fit the parameters of a dynamical ECG model to the individual user. This model captures the characteristic P, Q, R, S, and T waves as scaled Gaussians.
- Extended Kalman Filter (EKF): The EKF is employed for the actual denoising. The system uses the fitted dynamical ECG model for state prediction. Crucially, the in-ear heart sound timings are used as phase measurements for the EKF. This allows the filter to accurately track the cardiac cycle phase, even when the ECG amplitude signal is heavily corrupted by noise. The EKF's explainable nature allows for understanding how measurements influence state estimates.
Datasets
The study collected data from 18 healthy participants. For each participant, ear-ECG data was recorded using the OpenEarable ExG system, along with synchronized in-ear heart sounds. Ground truth Lead I ECG data was simultaneously recorded using a MAX30001G evaluation board with electrodes placed on the wrists. A total of 54 minutes of ear-ECG data was collected.
Key Insights and Ideas
- Heart Sounds as Precise Phase Anchors: The most significant insight is the effective use of in-ear heart sounds not just for timing, but as a low-noise phase measurement for the EKF. This overcomes the challenge of low SNR in the ECG amplitude signal itself.
- Explainable Denoising: By moving away from black-box autoencoders and towards an EKF with a dynamical model, the system offers greater explainability. This allows researchers and clinicians to understand why certain denoising decisions are made.
- Personalized ECG Models: Fitting a dynamical ECG model to each individual user's averaged signal allows the system to adapt to unique cardiac morphologies, leading to more accurate denoising.
- Earables as Viable ECG Platforms: The work demonstrates that ear-worn devices, with appropriate signal processing, can be a practical platform for continuous, non-invasive ECG monitoring.
Key Results
The system demonstrated significant improvements in ear-ECG signal quality:
- Mean Input SNR: -14.9 dB
- Mean Output SNR: 6.4 dB
- Mean SNR Improvement: 21.3 dB
- Mean Absolute Heart Rate Error: 3 BPM
- Mean Absolute QT Interval Error: 54 ms
These results indicate that the system not only cleans the signal but also preserves crucial diagnostic information like heart rate and QT intervals.
Limitations and Future Work
The authors acknowledge that the current system is evaluated for stationary participants. Future work will focus on deploying the system in clinical conditions to assess its performance on abnormal ECG morphologies. They anticipate that straightforward morphologies like inverted T-waves should be handled well, but conditions with significant temporal variations, such as premature ventricular contractions and fibrillation, will require further adaptations to the system.
Summary updated