Dynamic change of corneal thickness before and after SBK using the Pentacam measurement system
Lists
Only you see this
Abstract
Objective To compare the agreement and repeatability of the Pentacam with ultrasound pachymetry in the measurement of central corneal thickness (CCT). Methods A total of 130 patients (130 eyes) underdoing Sub-Bowman-Keratomileusis from March 2014 to October 2015 were included. Sequential measurements of CCT were performed with Pentacam and A-mode ultrasound pachymetry. The cutting depth in surgery was used to analysis the consistency of two ways measuring central corneal thickness. Results The mean values of CCT were 471.0-601.0μm (552.4±28.3μm) and 474.0-590.0μm (547.1±27.7μm) using the Pentacam and the ultrasonic pachymetry, respectively. There was no statistically difference between the two instrument (t=1.98, P=0.31). Mean values of CCT in 3 months follow-up after SBK were 471.0-601.0μm (552.4±28.3μm) and 474.0-590.0μm (547.1±27.7μm), there was no statistically difference either (t=1.25, P=0.65>0.05). Conclusions The agreement between the Pentacam and the ultrasonic pachymetry is good, and the repeatability is also good for Pentacam which is a promising instrument for the evaluation before SBK. Key words: SBK; Central corneal thickness; Pentacam; A-mode ultrasound pachymetry