Abstract
In a diverse cohort of pediatric congenital heart disease (CHD), measurement of atrioventricular valve regurgitation (AVVR) using the direct jet measurement (JET) method is accurate and reliable compared to the volumetric (VOL) standard. Furthermore, JET serves as a superior method of internal validation compared to the annular inflow method (AIM). This study supports the use of 4D flow cardiovascular magnetic resonance (CMR) for the measurement of AVVR, potentially removing the need for expert prospective prescription during image acquisition required by 2D CMR.
Introduction
Atrioventricular valve regurgitation (AVVR) is a significant contributor to morbidity and mortality in children with congenital heart disease (CHD). Accurate quantification of this leak is crucial for clinical decision-making. The current clinical standard, the Volumetric Method (VOL) using 2D CMR, is often limited by its reliance on multiple measurements and the need for precise, expert-led alignment of imaging planes during the scan. 4D flow CMR offers a promising alternative by capturing a complete volume of blood flow data, allowing for retrospective analysis. This study evaluates the accuracy and reliability of direct jet measurement (JET) within 4D flow datasets compared to traditional methods.
Methods
The researchers conducted a retrospective study involving 44 pediatric patients with various CHD conditions, including Ebstein anomaly and atrioventricular septal defects. Four methods for calculating Regurgitant Volume (RVol) were compared:
- VOL (Clinical Standard): 2D ventricular stroke volume minus 4D semilunar valve forward flow.
- AIM_stat (Annular Inflow Method): Flow measurement across the valve using a static plane.
- AIM_track: Flow measurement using a plane that tracks valve motion.
- JET (Direct Measurement): Direct quantification of the regurgitant jet volume within the 4D flow data.
The study utilized cloud-based post-processing software to manage the computational demands of 4D flow analysis.
Results
The JET method demonstrated a strong correlation with the clinical standard (VOL) ($\rho = 0.73, p < 0.001$) and high accuracy, with a median difference of only -1.5%. Reliability was excellent for all methods, with intra-class correlation coefficients (ICC) between 0.94 and 0.99. Notably, the Annular Inflow Method (AIM) consistently underestimated the regurgitant volume compared to both VOL and JET. The post-processing time for the JET method was approximately 3.1 minutes, making it highly feasible for clinical use.
Discussion
The JET method provides a "one-stop shop" for measuring valve leaks, as it relies on a single direct measurement rather than the subtraction of multiple flow values. This reduces the potential for compounded errors. Additionally, 4D flow acquisition simplifies the scanning process, as it does not require the expert prospective alignment needed for 2D CMR. While technical challenges such as velocity aliasing were noted in 52% of cases, the researchers were still able to obtain accurate measurements by adjusting the analysis planes.
Conclusion
Direct measurement of regurgitant jets using 4D flow CMR is an accurate, reliable, and efficient method for quantifying AVVR in pediatric patients with complex CHD. This approach simplifies the clinical workflow and provides a robust alternative to traditional 2D CMR methods. Future research should focus on longitudinal studies to determine the prognostic value of these 4D measurements in predicting clinical outcomes.