Development of an image-guided non-vitrectomy subretinal access approach for trans-scleral cell and gene therapy delivery

Mandeep S. Singh, Shoujing Guo, Christopher B. Toomey, Minda M. McNally, Sarah Harris‐Bookman, Kang Li, Dzhalal Agakishiev, Shuwen Wei, Soohyun Lee, Kathleen Perrino, Kevin C. Wolfe, Jin U. KangbioRxiv 2025 · DOI

TL;DR

This paper introduces a novel, image-guided surgical approach (MISA) for delivering therapies to the subretinal space in the eye. By using swept-source OCT for real-time imaging through the sclera, MISA aims to bypass the morbidities associated with traditional vitrectomy-based methods, potentially improving outcomes for degenerative retinal diseases.

What Problem Does This Paper Address?

Delivering therapeutic agents, such as stem cells or gene therapy vectors, to the subretinal space is crucial for treating various degenerative retinal diseases like age-related macular degeneration and retinitis pigmentosa. However, current methods typically require a transvitreal approach, involving vitrectomy (removal of the vitreous gel) and a retinotomy (a surgical hole in the retina). These procedures carry significant risks, including cataract formation, retinal detachment, and intraocular hemorrhage, which limit the safety and efficacy of subretinal delivery. The authors aim to develop a less invasive method that avoids these complications.

Methods

The core innovation is the Minimally Invasive Subretinal Access (MISA) system, which combines three key components:

  • Common-Path Swept-Source Optical Coherence Tomography (CP-SSOCT): This imaging technology provides real-time, depth-resolved visualization of retinal structures from an ab externo (outside the eye) perspective. It's integrated into a needle to guide its insertion.
  • Coaxial Guide (COG): A device that stabilizes the CP-SSOCT-enabled needle on the scleral surface, ensuring precise and controlled advancement. It uses vacuum suction for secure attachment.
  • Subretinal Access Cannula (SAC): A specialized cannula designed to create a smooth passageway for therapeutic materials into the subretinal space after initial access.

The MISA procedure was pilot-tested in ex vivo bovine eyes and in vivo porcine ocular surgery. The process involved two phases: Phase 1 focused on penetrating the sclera and creating a subretinal bleb using the CP-SSOCT-guided needle, while Phase 2 involved inserting the SAC into this bleb and advancing it to the posterior pole.

Datasets

Not discussed in the paper. The study primarily uses ex vivo bovine eyes and in vivo porcine eyes for testing.

Key Insights and Ideas

  • "Ab Externo" Imaging is Key: The ability to image and guide instruments through the sclera, rather than requiring internal access, is a fundamental shift that bypasses many risks of traditional surgery.
  • OCT as a Real-Time Surgical Navigator: Integrating OCT directly into the surgical needle allows for immediate feedback on tissue depth and location, crucial for navigating delicate retinal layers.
  • Minimally Invasive Access is Feasible: The study demonstrates that it's possible to access the subretinal space without vitrectomy or retinotomy, a significant step towards safer retinal therapies.
  • Complications Highlight Need for Refinement: While promising, the encountered complications (retinal incarceration/perforation) underscore that further optimization is needed before clinical translation.

Key Results

  • CP-SSOCT imaging from the scleral side successfully identified retinal layers and the subretinal space in both ex vivo and in vivo animal models.
  • The MISA approach enabled placement of the subretinal access cannula in the posterior pole of porcine eyes.
  • Major complications observed included retinal incarceration (n=2) and retinal perforation (n=3) at the ocular entry site.
  • Minor complications included RPE displacement (n=4), mild vitreous hemorrhage (n=3), mild subretinal hemorrhage (n=1), and moderate sclerochoroidotomy site hemorrhage (n=1).
  • Mitigation strategies, such as reducing intraocular pressure and turning off infusion before choroidotomy, were developed and proved effective in preventing major complications in later trials.

Limitations and Future Work

The authors acknowledge that the observed major and minor complications currently limit the clinical implementation of the MISA approach. Priorities for future research include:

  • Improving the graphical user interface of the CP-SSOCT system.
  • Adjusting COG parameters to optimize the angle of subretinal approach.
  • Further technical and procedural refinements to enhance safety and efficacy.

The authors believe that with these improvements, the MISA system could be clinically implemented to deliver subretinal therapeutics without the need for vitrectomy or retinotomy, thereby reducing surgical morbidity and potentially improving clinical outcomes for cell and gene therapies in the eye.

Summary updated

Abstract

Abstract Purpose Regenerative therapies for retinal diseases include cell and gene therapy modalities that are targeted to the subretinal space. Several recent clinical trials have shown that the morbidity of surgical access is the major limitation of safe subretinal space delivery. We aimed to develop an image-guided procedure for minimally invasive subretinal access (MISA) as a platform to deliver therapeutic agents for the treatment of degenerative retinal diseases. Methods We engineered prototypes of a novel common-path swept source optical coherence tomography (CP-SSOCT)-enabled needle, coaxial guide (COG), and subretinal access cannula (SAC). We pilot tested the MISA procedure in ex vivo bovine eyes and in vivo porcine ocular surgery. Results A- and M-mode scan recordings of ex vivo and in vivo animal eye models demonstrated that CP-SSOCT imaging from the scleral side ( ab externo ) was capable of identifying the retinal laminae and the sub-retinal space. We show results from in vivo porcine MISA surgeries (N=4) using the novel CP-SSOCT-enabled sub-retinal injection needle, COG, and SAC through the transscleral approach. The MISA approach enabled subretinal device placement in the posterior pole, however, cases of retinal incarceration and retinal perforation were encountered. Conclusions We describe a novel CP-SSOCT-guided subretinal access approach that, with further optimization, may be useful in regenerative retinal surgery.
BibTeX
@misc{singh2025development,
  author = {Mandeep S. Singh and Shoujing Guo and Christopher B. Toomey and Minda M. McNally and Sarah Harris‐Bookman and Kang Li and Dzhalal Agakishiev and Shuwen Wei and Soohyun Lee and Kathleen Perrino and Kevin C. Wolfe and Jin U. Kang},
  title = {{Development of an image-guided non-vitrectomy subretinal access approach for trans-scleral cell and gene therapy delivery}},
  year = {2025},
  doi = {10.1101/2025.02.25.640217},
  howpublished = {bioRxiv},
}

Loading related content...

Figure