Development of a Smartphone-Based Fluorescent Immunochromatographic Assay Strip Reader

Qi Zheng, Huihuang Wu, Haiyan Jiang, Jiejie Yang, Yueming GaoSensors 2020 · DOI

TL;DR

This paper presents a novel smartphone-based reader for fluorescence immunochromatographic assays (FICA), leveraging a smartphone's built-in camera and processing power to create a portable, low-cost, and versatile diagnostic tool. The system achieves quantitative detection of analytes like HCG and Aflatoxin B1 with high accuracy, addressing the cost and complexity limitations of existing FICA readers.

What Problem Does This Paper Address?

Fluorescence immunochromatographic assays (FICA) are valuable for rapid, sensitive, and precise point-of-care testing (POCT). However, existing FICA strip readers often rely on expensive, high-precision CMOS image sensors or complex photoelectric scanning mechanisms. This increases their cost, reduces portability, and can make them difficult to operate and maintain, hindering widespread adoption in resource-limited settings or for broad field applications. The authors aim to overcome these limitations by developing a more accessible FICA reader using readily available smartphone technology.

Methods

The core of the proposed system is a smartphone-based reader that integrates an optical module with an Android smartphone.

  • Optical Module: This module is designed to provide uniform illumination and capture clear images of the test strips. It features a single LED excitation light source with a reflective surface (aluminum foil tape) to expand illumination. Crucially, the excitation light source and light filter are designed to be interchangeable, allowing compatibility with test strips using different fluorescent markers. A macro lens is incorporated to reduce the minimum imaging distance, enabling a more compact optical module. The module is constructed from light-cured plastic and designed to be light-tight to ensure consistent testing conditions.
  • Image Acquisition and Processing: An Android smartphone captures images of the test strip within the optical module. Image denoising is performed to mitigate noise from environmental factors.
  • Line Recognition: Instead of more computationally intensive algorithms, a sliding window algorithm is employed to efficiently identify the test (T) and control (C) lines on the strip image. This algorithm traverses the image with small steps, minimizing system resource usage and improving processing speed on the smartphone.
  • Characteristic Value Calculation: The gray values of the T and C lines are extracted. The characteristic value of the strip is calculated as the ratio of the gray value of the T line to the C line, with background gray value subtraction to improve accuracy. The specific formula depends on whether a double-antibody sandwich ICA (positive correlation) or competitive ICA (negative correlation) is used.

Datasets

The paper does not describe the creation of new datasets for training. Instead, it uses standard FICA test strips for performance verification.

  • Human Chorionic Gonadotrophin (HCG) test strips: Used for quantitative detection.
  • Aflatoxin B1 (AFB1) test strips: Also used for quantitative detection.

The experiments involve preparing solutions of known concentrations of HCG and AFB1 to create calibration curves and test the accuracy of the developed reader.

Key Insights and Ideas

  • Leveraging Smartphone Ubiquity: The paper highlights the immense potential of smartphones as powerful, portable, and cost-effective platforms for advanced diagnostics, capitalizing on their integrated CMOS sensors and processing capabilities.
  • Modular Optical Design: The interchangeable excitation light source and filter are key to the system's versatility, allowing it to adapt to various fluorescent markers used in different FICA strips.
  • Efficient Sliding Window Algorithm: Employing a sliding window algorithm for line detection on a smartphone is a clever optimization that balances accuracy with computational efficiency, crucial for mobile applications.
  • Cost-Effective POCT: By replacing expensive components with smartphone technology and a simplified optical module, the system significantly reduces the cost of FICA readers, making advanced diagnostics more accessible.

Key Results

The smartphone-based FICA strip reader demonstrated strong performance in quantitative detection:

  • Human Chorionic Gonadotrophin (HCG):
    • Linear detection range: 10 to 5000 mIU/mL (R² = 0.95).
  • Maximum relative error: Less than 9.41%.
  • Limit of Detection (LoD): 4.7 mIU/mL.
  • Aflatoxin B1 (AFB1):
    • Linear detection range: 5 to 4000 pg/mL (R² = 0.99).
  • Maximum relative error: Less than 12.71%.
  • Limit of Detection (LoD): 3.9 pg/mL.

These results indicate high portability, versatility, and accuracy, comparable to or exceeding some existing methods.

Limitations and Future Work

  • Limitations:
    • The paper acknowledges that while the sliding window algorithm is efficient, the accuracy of line detection can be influenced by the width of the window relative to the T and C lines.
  • The accuracy tests show a maximum relative error of up to 9.41% for HCG and 12.71% for AFB1, indicating room for further improvement in precision.
  • The optical module's performance is dependent on the quality of the smartphone camera and its ability to focus accurately.
  • Future Work:
    • The authors suggest that the smartphone-based FICA strip reader has broad application prospects in POCT due to its inherent advantages.
  • Further optimization of the optical module and image processing algorithms could lead to even higher accuracy and lower detection limits.
  • Expanding the range of detectable analytes by developing more versatile optical modules and assay protocols would be a logical next step.

Summary updated

Abstract

Fluorescence immunochromatographic assay (FICA) is a rapid immunoassay technique that has the characteristics of high precision and sensitivity. Although image FICA strip readers have the advantages of high portability and easy operation, the use of high-precision complementary metal oxide semiconductor (CMOS) image sensors leads to an increase in overall cost. Considering the popularity of CMOS image sensors in smartphones and their powerful processing functions, this work developed a smartphone-based FICA strip reader. An optical module suitable for the test strips with different fluorescent markers was designed by replacing the excitation light source and the light filter. An android smartphone was used for image acquisition and image denoising. Then, the test and control lines of the test strip image were recognized by the sliding window algorithm. Finally, the characteristic value of the strip image was calculated. A linear detection range from 10 to 5000 mIU/mL (<i>R</i><sup>2</sup> = 0.95) was obtained for human chorionic gonadotrophin with the maximum relative error less than 9.41%, and a linear detection range from 5 to 4000 pg/mL (<i>R</i><sup>2</sup> = 0.99) was obtained for aflatoxin B1, with the maximum relative error less than 12.71%. Therefore, the smartphone-based FICA strip reader had high portability, versatility, and accuracy.
BibTeX
@inproceedings{zheng2020development,
  author = {Qi Zheng and Huihuang Wu and Haiyan Jiang and Jiejie Yang and Yueming Gao},
  title = {{Development of a Smartphone-Based Fluorescent Immunochromatographic Assay Strip Reader}},
  booktitle = {Sensors},
  year = {2020},
  doi = {10.3390/s20164521},
}
Figure