Development of a Highly Sensitive Serum Neurofilament Light Chain Assay on an Automated Immunoassay Platform

Stephen Lee, Tatiana Plavina, Carol M. Singh, Kuangnan Xiong, Xiaolei Qiu, Richard A. Rudick, Peter A. Calabresi, Lauren Stevenson, Danielle Graham, Denitza Raitcheva, Christopher Green, Madeleine Matias, Arejas J. UzgirismedRxiv 2022 · DOI

TL;DR

This paper details the development and validation of a highly sensitive assay for Neurofilament light chain (NfL)—a critical biomarker for axonal damage—on standard automated clinical immunoassay platforms. By achieving a lower limit of quantitation (LLoQ) of 1.85 pg/mL, the researchers demonstrate that this vital neurological marker can be moved from specialized research labs into routine clinical practice.

What Problem Does This Paper Address?

Neurofilament light chain (NfL) is a protein found in the "skeleton" of neurons. When brain cells are damaged or die due to conditions like Multiple Sclerosis (MS) or Amyotrophic Lateral Sclerosis (ALS), NfL leaks into the cerebrospinal fluid (CSF) and eventually into the blood.

While NfL is a fantastic "check engine light" for the brain, measuring it in blood is incredibly difficult because the concentrations are minute—often in the single-digit picograms per milliliter (pg/mL). Until now, measuring blood-based NfL required specialized, low-throughput "research use only" platforms. To make NfL a standard tool for doctors to monitor disease progression or drug effectiveness, we need an assay that is sensitive, standardized, and compatible with the high-throughput automated machines already found in hospital laboratories.

Methods

The researchers developed an immunometric assay designed to run on the Siemens ADVIA Centaur XP and Atellica Solution platforms.

  • The Mechanism: The assay uses a "sandwich" approach. One monoclonal antibody is attached to a magnetic bead to capture the NfL, while a second antibody is labeled with an acridinium ester (AE). When triggered, the AE emits light (chemiluminescence) proportional to the amount of NfL present.
  • Validation: The team followed rigorous Clinical and Laboratory Standards Institute (CLSI) protocols to test sensitivity, precision, linearity, and potential interference from common substances (like bilirubin or biotin).
  • Optimization: While initially developed for serum, the assay was optimized and validated for plasma and CSF to ensure "specimen equivalence."

Datasets

The study utilized several distinct datasets for validation:

  • Analytical Samples: Contrived samples using NfL-depleted serum spiked with recombinant human NfL to establish detection limits.
  • MS Cohort: 458 clinical serum samples from the MSPATHS biorepository and the ADVANCE clinical trial (evaluating peginterferon beta-1a).
  • ALS Cohort: CSF samples from healthy controls and patients with a definite ALS diagnosis.
  • Method Comparison: A subset of samples was run in parallel on the Quanterix SIMOA platform (the current research gold standard) to ensure results were comparable.

Key Insights and Ideas

  • Democratizing Diagnostics: The most significant "aha" moment is that sub-2 pg/mL sensitivity is achievable on a routine clinical platform. This bridges the gap between high-end proteomics research and daily patient care.
  • High Correlation: The assay showed a high correlation ($R^2 = 0.907$) with the SIMOA platform, meaning existing research findings can likely be translated to this new, more accessible platform.
  • Clinical Mirroring: Serum NfL levels weren't just numbers; they correlated significantly with physical evidence of disease. In MS patients, higher sNfL levels at baseline predicted a higher number of new T2 lesions on MRI six months later.
  • Speed and Scale: Using the Atellica platform, the time to the first result is just 51 minutes, with a throughput of 171 samples per hour—orders of magnitude faster than manual research methods.

Key Results

  • Sensitivity: A Lower Limit of Quantitation (LLoQ) of 1.85 pg/mL in serum.
  • Precision: Excellent repeatability with a coefficient of variation (CV) of <6%.
  • Linearity: The assay is linear across a wide range (1–646 pg/mL), which is crucial for capturing both healthy baselines and acute spikes in damage.
  • Specificity: Very low cross-reactivity (<0.7%) with related proteins (NfM and NfH), ensuring the signal is truly from the "light" chain.
  • Clinical Association: In ALS patients, CSF NfL levels were roughly two orders of magnitude higher than in healthy controls, providing a clear diagnostic signal.

Limitations and Future Work

The authors acknowledge that while the assay is analytically robust, clinical implementation requires normative reference ranges. NfL levels naturally increase as people age, so "normal" must be defined for different age brackets before a single test result can be used for diagnosis.

Future work will likely focus on establishing these age-adjusted baselines and testing the assay's utility in other neurodegenerative conditions, such as Alzheimer’s or Parkinson’s disease. Additionally, the researchers suggest that this assay could be a powerful tool in clinical trials to provide a "sensitive measurement of treatment effect," potentially reducing the sample sizes needed for early-stage drug testing.

Summary updated

Abstract

ABSTRACT Background Neurofilament light chain (NfL) is an axonal cytoskeletal protein that is released into the extracellular space following neuronal or axonal injury associated with neurological conditions such as multiple sclerosis (MS), amyotrophic lateral sclerosis (ALS), and other diseases. NfL is detectable in cerebrospinal fluid (CSF) and blood. Numerous studies in MS have demonstrated that NfL correlates with disease activity, predicts disease progression, and is reduced by treatment with MS disease-modifying drugs, making NfL an attractive candidate to supplement existing clinical and imaging measures in MS. However, for NfL to achieve its potential as a clinically useful biomarker for clinical decision-making or drug development, a standardized, practical, widely accessible assay is needed. Our objective was to validate the analytical performance of the novel serum neurofilament light (sNfl) assay on the ADVIA Centaur® XP immunoassay system. Methods The research assay was evaluated on the ADVIA Centaur XP immunoassay system from Siemens Healthineers. The lower limit of quantitation (LLoQ), intra-assay variation, assay range, cross-reactivity with neurofilament medium and heavy chains, and effect of interfering substances were determined. NfL assay values in serum and CSF were compared with radiological and clinical disease activity measures in patients with MS and ALS, respectively. This assay was further optimized to utilize serum, plasma, and CSF sample types and transferred to Siemens’ CLIA laboratory, where it was analytically validated as a laboratory-developed test. Results In this study, a LLoQ of 1.85 pg/mL, intra-assay variation of &lt;6%, and an assay range of up to 646 pg/mL were demonstrated. A cross-reactivity of &lt;0.7% with neurofilament medium and heavy chains was observed, and the assay was not significantly affected by various interfering substances encountered in clinical specimens. Serum and CSF NfL assay values were associated with radiological and clinical disease activity measures in patients with MS and ALS, respectively. Conclusion The analytical performance of the NfL assay fulfilled all acceptance criteria; therefore, we believe the assay is acceptable for use in both research and clinical practice settings to determine elevated sNfL levels in patients.