Development of a Blood Pressure Measurement Instrument with Active Cuff Pressure Control Schemes

Chung‐Hsien Kuo, Chun-Ju Wu, Hung-Chyun Chou, Guanting Chen, Yu-Cheng KuoJournal of Healthcare Engineering 2017 · DOI

TL;DR

This paper introduces a novel oscillometric blood pressure (BP) measurement device that uses a variable volume chamber controlled by a screw motor to actively manage cuff pressure. This approach significantly reduces air turbulence and measurement time, offering two modes (AIM and ADM) for faster and more consistent BP readings compared to conventional devices.

What Problem Does This Paper Address?

Regular blood pressure monitoring is crucial for managing chronic diseases like hypertension. Traditional oscillometric BP monitors, while convenient and non-invasive, often suffer from long measurement times and potential inaccuracies due to air turbulence generated by standard air pumps during the inflation phase. This turbulence can interfere with the detection of key BP characteristics (systolic and diastolic pressures), forcing measurements to occur during the slower deflation cycle. The paper aims to overcome these limitations by developing a more efficient and precise BP measurement instrument.

Methods

The core innovation is a variable volume chamber that replaces the conventional air pump. This chamber is driven by a single-acting cylinder controlled by a DC motor and a linear screw mechanism. This setup allows for precise, closed-loop control of cuff pressure.

Key methodological aspects include:

  • Active Pressure Control: A PID controller manages the DC motor to achieve desired pressure trajectories within the cuff. This allows for active pressure adjustments during both inflation and deflation.
  • Nonlinear Trajectory Planning: Instead of a simple linear inflation/deflation, the system employs carefully designed pressure-trajectory functions (sigmoid, cosine, and third-order polynomial) to optimize different phases of the measurement. This is crucial for both speed and accuracy.
  • Two Measurement Modes:
    • AIM (Air Injection Measurement): This mode aims for speed by measuring BP characteristics during the inflation phase, leveraging the stable air injection from the new chamber.
  • ADM (Accurate Dual-Way Measurement): This mode uses a multi-stage approach: a fast initial inflation to get rough BP estimates, followed by a more precise, variable-resolution deflation phase guided by the initial estimates. This balances speed with improved consistency.
  • Signal Processing: Standard techniques are used to process pressure signals from a sensor, including filtering to separate direct (cuff pressure) and alternating (oscillometric pulse) signals.

Datasets

Not discussed in the paper. The evaluation relies on experiments conducted with healthy human subjects.

Key Insights and Ideas

  • Eliminating Air Turbulence is Key: The paper highlights that air turbulence from traditional pumps is a major source of noise and inaccuracy in oscillometric BP measurements. The variable volume chamber effectively mitigates this.
  • Active Control for Efficiency: By actively controlling the pressure trajectory, the system can perform measurements during inflation (AIM mode) and optimize deflation (ADM mode), significantly reducing overall measurement time.
  • Tailored Trajectories for Different Phases: The use of different mathematical functions (sigmoid, cosine, polynomial) for distinct stages of the pressure cycle is a clever way to balance motor vibration reduction, fast pressure changes, and precise detection of BP characteristics.
  • ADM Mode Balances Speed and Consistency: The ADM mode's multi-stage approach, using rough estimates to guide precise measurements, offers a compelling trade-off between speed and improved measurement consistency.

Key Results

  • Reduced Measurement Time:
    • AIM mode reduced measurement time by 34.21% compared to a commercial BP monitor.
  • ADM mode reduced measurement time by 15.78% compared to a commercial BP monitor.
  • Improved Consistency: The ADM mode demonstrated more consistent measurements (lower standard deviation) compared to a commercial BP monitor.
  • Feasibility Demonstrated: The system successfully measured SBP and DBP using both AIM and ADM modes in healthy subjects. The experimental setup, though using homemade circuits, showed promising results.

Limitations and Future Work

  • Accuracy Comparison Caveat: The authors note that the experiment setting was not primarily for BP accuracy comparison, as their homemade circuits and mechanical designs were not certified. Bias and interference were acknowledged as potential concerns.
  • Future Work:
    • Optimizing control parameters based on characteristic models of the cylinder and cuff for further improvement in stability and efficiency.
  • Conducting clinical feasibility studies with real patients under IRB approval.
  • Verifying the system against IEC EN 1060-4 standards.
  • Exploring further improvements to AIM accuracy and ADM speed through refined trajectory planning.

Summary updated

Abstract

This paper presents an oscillometric blood pressure (BP) measurement approach based on the active control schemes of cuff pressure. Compared with conventional electronic BP instruments, the novelty of the proposed BP measurement approach is to utilize a variable volume chamber which actively and stably alters the cuff pressure during inflating or deflating cycles. The variable volume chamber is operated with a closed-loop pressure control scheme, and it is activated by controlling the piston position of a single-acting cylinder driven by a screw motor. Therefore, the variable volume chamber could significantly eliminate the air turbulence disturbance during the air injection stage when compared to an air pump mechanism. Furthermore, the proposed active BP measurement approach is capable of measuring BP characteristics, including systolic blood pressure (SBP) and diastolic blood pressure (DBP), during the inflating cycle. Two modes of air injection measurement (AIM) and accurate dual-way measurement (ADM) were proposed. According to the healthy subject experiment results, AIM reduced 34.21% and ADM reduced 15.78% of the measurement time when compared to a commercial BP monitor. Furthermore, the ADM performed much consistently (i.e., less standard deviation) in the measurements when compared to a commercial BP monitor.
BibTeX
@article{kuo2017development,
  author = {Chung‐Hsien Kuo and Chun-Ju Wu and Hung-Chyun Chou and Guanting Chen and Yu-Cheng Kuo},
  title = {{Development of a Blood Pressure Measurement Instrument with Active Cuff Pressure Control Schemes}},
  journal = {Journal of Healthcare Engineering},
  year = {2017},
  doi = {10.1155/2017/9128745},
}
Figure